摘要
目的及时、方便、准确地监测烧伤患者的血糖变化,为正确的诊断、治疗提供依据。方法使用同一血糖仪同时分别检测同一患者的静脉血糖值与末梢血糖值,共采集了71例烧伤患者和50例门诊患者的血糖数据,进行自身对照及相关分析。结果静脉血糖值与末梢血糖值自身对照差异有显著性(P<0.01);两者之间呈直线相关,烧伤患者的相关系数r值=0.939,门诊患者的相关系数r值=0.916;烧伤患者的回归方程:Y=0.889+0.959X,门诊患者的回归方程为:Y=0.211+1.069X。结论将血糖仪检测的患者末梢血糖结果代入上述方程,换算成静脉血糖值,判断血糖水平较准确。
Objective To monitor the blood glucose concentration, and then provide the scientific evidence of diagnosis and treatment for burn patients in time. Methods Using the same glucose meter to monitor the intravenous glucose and the peripheral glucose respectively for the same patient in the same time, and then compared the difference of them. The bload glucose samples were taken from 71 burn patients and 50 outpatients. Results There were significant difference between the intravenous glucose and the peripheral glucose, P〈0.01. There were linear correlation between the intravenous gluccse and the peripheral glucose not only in the 71 burn patients, but also in the 50 outpatients, r = 0. 939 and 0.916 respectively, the liner equations were Y= 0. 889 + 0. 959X and Y=0.211 + 1.069X respectively. Conclusion The peripheral bload glucose value can be changed into the intravenous bload glucose value by the linear equation, which can show the blood - sugar concentration more exactly.
关键词
静脉血糖
末梢血糖
烧伤
Intravenous blood glucose
Peripheral blood glucose
Burn