Si(Li)探测器低能区的效率刻度 被引量:2

Efficiency Calibration of Si(Li) Detector in Low-Energy Region
在线阅读 下载PDF
导出
摘要 通过测量19 keV电子束轰击厚碳靶产生的韧致辐射能谱,在低能区(>1 keV)对Si(Li)探测器进行了探测效率刻度。厚碳靶的理论韧致辐射能谱由Monte-Carlo程序PENELOPE计算,并用241Am标准放射源确定出效率刻度曲线的绝对值。采用本工作的刻度方法确定的效率刻度曲线误差主要来源于用标准放射源绝对化的误差,约为5%。将所得初步实验结果与理论计算值进行了比较,并采用最小二乘法对探测器各厚度参数进行了拟合,除Au接触层厚度外,拟合的各厚度值与探测器生产商提供的数据较为吻合。 The detection efficiency calibration of Si(Li) detector in low-energy region down to 1 keV was carried out by measuring the bremsstrahlung spectrum of thick-carbon-target bombarded by 19 keV electron beam. The coupled electron-photon transport Monte-Carlo computer code PENELOPE was employed for theoretical calculation of thick-carbon-target bremsstrahlung spectrum. The absolute value of the efficiency calibration was determined from the use of 241 Am radioactive standard source. The uncertainty for the efficiency calibration with present method is mainly stemmed from the use of ^241Am radioactive standard source and it was estimated to be about 5%. The preliminary experimental efficiency calibration values were also compared with calculation values based on a simple model. The detector thickness parameters were also determined by a nonlinear least-squares fit. Except the Au-contact layer, the other thickness parameters are in acceptable agreement with that provided by detector manufacturer.
出处 《原子能科学技术》 EI CAS CSCD 北大核心 2006年第1期83-87,共5页 Atomic Energy Science and Technology
基金 国家自然科学基金与中国工程物理研究院联合基金资助项目(10276029) 高等学校博士学科点专项科研基金资助项目(20020610019) 国际原子能机构资助项目(12354/R0/RBF)
关键词 探测效率 Si(Li)探测器 MONTE-CARLO方法 detection efficiency Si(Li) detector Monte-Carlo method
  • 相关文献

参考文献26

  • 1AGOSTINELLI S, ALLISON J, AMAKO K, et al. GEANT4 A simulation toolkit[J]. Nucl Instrum Methods, 2003, A506(3):250-303.
  • 2HANSEN J S, McGEORGE J C, NIX D, et al.Accurate efficiency calibration and properties of semiconductor detectors for low-energy photons[J]. Nucl Instrum Methods, 1973, 106 (2):365-379.
  • 3GALLAGHER W J, CIPOLLA S J. A modelbased efficiency calibration of a Si(Li) detector in the energy region from 3 to 140 keV[J]. Nucl Instrum Methods, 1974,122(3):405-414.
  • 4CAMPBELL J L, McGHEE P L. State-of-theart efficiency determination for a Si(Li) X ray detector in the 3-40 keV energy range[J]. Nuel Instrum Methods, 1986,A248(2/3): 393-404.
  • 5LENNARD W N, PHILLIPS D. Absolute measurement of the photopeak efficiency for a Si(Li) detector: 0. 52 8. 04 keV[J]. Nuel Instrum Methods, 1979,166(3):521-529.
  • 6PAJEK M, KOBZEV A P, SANDRIK R, et al.Accurate efficiency determination of a Si(Li) detector in the Si-K and AuM absorption edge energy region[J]. Nucl Instrum Methods, 1989,B42(3);346-358.
  • 7TRIBEDI L C, TANDON P N. Efficiency determination of a Si(Li) X-ray detector in the energy range 1-20 keV[J]. Nucl Instrum Methods,1992,B69(2/3):178-181.
  • 8UZONYI I, SZABO Gy, BORBELY KISS I, et al. Calibration of an UTW Si(Li) detector in the 0.28-22. 1 keV energy region at a nuclear microprobe[J]. Nucl Instrum Methods, 2003,B210:147-152.
  • 9WEATHERS D L, DUGGAN J L, McNEIR M R, et al. Efficiency determination for a windowless Si(Li) X-ray detector for photon energies below 5 keV using atomic-field bremsstrahlung[J].Nucl Instrum Methods, 1991, B56/57 (2):964-967.
  • 10TSCHISCHGALE J, KUCHLER D, LEHNERT U, et al. Efficiency calibration of a Si(Li) detector in the low-energy region using electron bremsstralung [J]. Nucl Instrum Methods,1997, A400(2/3):387-391.

二级参考文献1

共引文献9

同被引文献11

引证文献2

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部