期刊文献+

液芯光纤长光路法在水溶液中测定Cr、Cu

Determination of(Ⅵ),Cu(Ⅱ)in Aqueous Solution by Liquid-Core Optical Fiber Spectrophotometry
在线阅读 下载PDF
导出
摘要 测量了低折射率液体(n>1.346)和水溶液液芯光纤的导光特性和光谱特性.研究了液芯光纤长光路法在70%乙醇-水溶液体系和50%二氧六环-水溶液体系中测定Cr(Ⅵ)、Cu(Ⅱ)的方法.液芯光纤长110cm时,测定Cr、Cu的灵敏度分别为0.052、0.017ng·mL-1;线性范围0~25、0~24ng/mL;RSD值为1.0%~4.8%.对国标GSBE50009-88样品和明胶中铬的测量值与标准值或AAS测量值吻合,回收率为94%~102%. This paper presents the transmission characteristics and spectrum for liquid-core optical fiber filled with low refractive index liquids(n>1.346)and aqueous solutions.The methods are studied for the determination of Cr(Ⅵ),Cu(Ⅱ) in aqueous solutions of about 70% alcohol and 50% dioxane.With the 110 cm liquid-core optical fiber in length,the sensitivities(1/ng·mL ̄(-1))of Cr(Ⅵ), Cu(Ⅱ)are 0. 052 and 0.017 respectively;the linear ranges are 0~25,0~24 ng/mL with RSD being 1.0%~4.8%.The Cr in the animal gum and national standard are determined with recoveries of 94%~102%.
出处 《高等学校化学学报》 SCIE EI CAS CSCD 北大核心 1996年第3期377-380,共4页 Chemical Journal of Chinese Universities
基金 国家自然科学基金
关键词 液芯光纤 水溶液 光度法 Liquid-core optical fiber,Aqueous solution, Cr(Ⅵ),Cu(Ⅱ)