期刊文献+

四苯硼根离子选择电极的研制及其应用 被引量:2

Development of Tetraphenylboron Ion-Selective Electrode and Its Application
在线阅读 下载PDF
导出
摘要 以三辛基十二烷基铵-四苯硼缔合物为电活性物质、邻苯二甲酸二异辛酯为溶剂介体研制了四苯硼根涂碳PVC膜离子选择电极,测试了它的性能,结果良好:以它为指示电极、四苯硼钠标准溶液为滴定剂,用电位滴定法测定了片剂中黄连素的含量,结果与药典法一致.本文还用该电极测试计算了四苯硼钾、四苯硼银和四苯硼黄连系的溶度积常数,前二者与文献一致,后者尚未见报导. The coated carbon PVC matrix membrane tetraphenylboron (TPB-) ion-se-lective electrode was prepared and developed with trioctylodoe cylammonium-TPB aselectroactive material and disisooctylphthalate as mediater solvent. The electrode char-acteristics were measured with good result. The content of berberine in its hydrochlo-ride tablets were determined by potentiometer titration using the TPB- electrode as indi-cation eletrode and NaTPB standard solution as titrant. The result obtained was in ac-cordance with the that of Chinese Pharmacopoeia method. The solubility product con-stants were measured and calculated for K-TPB, Ag-TPB and berbrine-TPB, the for-mer two values were well agreement with the values of the literature.
机构地区 山西大学化学系
出处 《传感技术学报》 CAS CSCD 1996年第1期37-40,共4页 Chinese Journal of Sensors and Actuators
基金 山西省自然科学基金 山西省回国留学人员科研基金资助课题
关键词 离子选择电极 四苯硼 PVC膜 电极 ion-selective electrode tetraphenylboron potentiometer tritration berbrine solutility product constant
  • 相关文献

参考文献9

  • 1蒋雅琴,第六届全国化学传感器学术交流会论文集,1994年
  • 2高鸿,示波药物分析,1992年
  • 3安登魁,药物分析(第3版),1992年
  • 4张国雄,化学学报,1990年,38卷,3期,224页
  • 5李为群,化学传感器,1988年,8卷,3期,19页
  • 6刘明辉,化学传感器,1986年,6卷,3期,63页
  • 7团体著者,中华人民共和国药典.1,1985年
  • 8俞汝勤,分析化学,1983年,11卷,5期,343页
  • 9张国雄,分析化学,1983年,11卷,7期,532页