摘要
笔者对医用观片灯的现状进行了调查,结果发现被检的66台灯中,最高亮度1305Lx,最低160Lx,平均760Lx,达不到国家卫生部的标准。亮度严重不足,直接影响了诊断质量,亟待解决。
After we investigated the condition of 66 viewboxs used in hospitals, we discovered that none of them are up to the standard. The maximum brightness of them is 1305 Lx, the minimum is 160LX, and the average is 760 Lx. It is far lower then the standard of Ministry of Health (the brightness of viewbox must not be lower than 1350 cd/m2, i.e. 1350 Lx). The brightness of the viewboxes is so insuff icient that it has directly had influence on the diagnosis quality.For this reason. it must be corrected imme diately.
出处
《中国医疗器械杂志》
CAS
1995年第5期282-283,297,共3页
Chinese Journal of Medical Instrumentation