期刊文献+

生物材料血液相容性综合评价体系的研究 被引量:15

STUDY ON COMPREHENSIVE EVALUATION OFHENOCOMPATIBILITY OF BIOMATERIALSBY MULTIPARAMETER TESTINGMETHODS
暂未订购
导出
摘要 本文提出由四个体外和半体内试验综合评价生物材料血液相容性的体系,同时用此综合评价体系评价了六种用于心血管系统的聚氨酯材料和一种硅橡胶材料的血液相容性。试验结果表明本文提出的四参数综合评价体系可较真实地评价生物材料对血液的急性反应,并可对生物材料的血液相容性进行粗筛。 <ABSTRACT> In this report, we describe four multiparameter testing methods for evaluating hemocompatibility of blomaterials and established comprehensive cvaluation system testing methods. These methods are: 1 ) kinetic clotting time in vitro test,2) measurement of fibrinogen adhesion ability on surface of testing materials with radio-labelling technique in canine AV-series shunt model ex vivo, 3 ) observation of platelel adhesion on surface of testing materials by SEM, and 4 ) dynamic thrombus formation test in vitro Six polyurethanes (Cardiothane, CUST-Ⅰ, ZH- Ⅲ, Biomer, CUST- Ⅱ and Unithane 80E ) and one silicone (DCGB ) were used as testing materials in this paper. Results of this experiment show that the comprehensive evaluating system of multiparameter tests can more exactiy evaluate me - compatibility of biomaturials, especially for evaluation of acute blood Inter- action.
出处 《中国生物医学工程学报》 CAS CSCD 北大核心 1989年第4期212-221,共10页 Chinese Journal of Biomedical Engineering
基金 卫生部科学研究基金资助课题
  • 相关文献

参考文献9

  • 1奚廷裴,第三届国际生物材料会议论文集,1988年
  • 2奚廷裴,医用物理和生物医学工程国际会议论文集,1988年
  • 3奚廷裴,中国生物医学工程学报,1987年,4期,245页
  • 4团体著者,1987年
  • 5钟银屏,生物医学工程杂志,1985年,1期,12页
  • 6团体著者,1985年
  • 7查晓,生物医学工程杂志,1984年,7期,35页
  • 8杨福良,高分子学报,1982年,4期,302页
  • 9黄祖--,科学通报,1982年,12期,725页