摘要
利用伴随粒子飞行时间方法,通过测量14.7MeV中子在聚乙烯和碳样品上的次级中子谱,得到了在0.99—14.7MeV能量范围内ST-451液体闪烁体中子探测器的探测效率值,其总误差<4.7%,其中统计误差<3.5%.
Using the associated particle TOF method,the efficiency curve of ST-451 liquid scintillation neutron detector has been obtained by measuring the spectra of secondary neutrons from polyethylene and carbon induced by 14.7MeV neutrons.The neutron energy for the efficiency measurement ranges from 0.9 to 14.7MeV. The statistical error is less than 3.5% and the total error less than 4.7%.
出处
《核技术》
CAS
CSCD
北大核心
1995年第2期104-107,共4页
Nuclear Techniques
关键词
液体闪烁体
中子探测器
效率
Liquid scintillator,Neutron detector,Efficiency