期刊文献+

流动注射光导纤维动力学荧光法测定锰——锰-硫胺素-高碘酸钠体系 被引量:4

Fluorescence Spectrophotometric Determination of Trace Manganese ( I ) by Optical Fiber Kinetic Method -Flow Injection Analysis With Manganese-Thiamine-Sodium Periodate System
在线阅读 下载PDF
导出
摘要 本文提出用硫胺素-高碘酸钠作为指示反应,流动注射光导纤维动力学荧光法测定痕量锰.检测限达 0.4 ng/mL;相对标准偏差小于 1%;采样频率为 45样次/h.用于直接测定纯铝及铝合金的痕量锰,结果满意. The fluosescence spectrophotometric determination of trace manganese ( Ⅱ ) by optical fiber kinetic method with FIA was developed using sodium periodate and thiamine as an indicating reaction. The instability of thiamine as a kinetic fluorescent indicator has been overcome. The detection limit of manganese ( Ⅱ ) is 4. 0×10-10 g/mL, and the linear relationship between the fluoresence intensity and the concentration of manganese ( Ⅱ ) is found in the range of 0. 0082-1. 0 ug/mL; the relative standard deviation is less than 1%, and the sampling speed is 45 samples/h. The proposed method can directly be applied to the determination of manganese in aluminum and aluminum alloy.
出处 《分析化学》 SCIE EI CAS CSCD 北大核心 1995年第3期299-301,共3页 Chinese Journal of Analytical Chemistry
基金 国家自然科学基金资助项目
关键词 催化动力学 流动注射分析 光导纤维光谱法 Manganese, catalytic kinetics, flow injection analysis, optical fiber spectrometry