三维适形放射治疗位置与剂量精度的验证 被引量:6

The Verification of the Accuracy of Location and Dose for 3DCRT
暂未订购
导出
摘要 目的:验证三维适形放射治疗位置与剂量精度。方法:采用胶片法和电离室法,对所使用的治疗计划系统和加速器距离精度、靶中心位置精度及等中心吸收剂量进行实测验证。结果:治疗计划系统所显示图像的距离精度<1%,靶中心位置精度<2mm,等中心吸收剂量精度<1%。结论:所验证的三维适形放疗位置与剂量精度符合临床要求,验证方法可行。 Objective: To validate the accuracy of location and dose for 3DCRT. Methods: In this paper, the distance accuracy and the target accuracy as well as the absorbed dose in isocenter for 2300 C/D linear accelerator and TPS were measured and verified by film and ionization chamber. Results: The distance accuracy of images displayed by 3D TPS was <1%, and the location accuracy in isocenter <2 mm. The absorbed dose accuracy in isocenter was <1%. Thus, the accuracy of location and dose verified for 3DCRT were consistent with the clinical requirement, and the verification methods were feasible.
出处 《中国医学物理学杂志》 CSCD 2004年第3期135-136,共2页 Chinese Journal of Medical Physics
关键词 适形放疗 位置精度 剂量精度 验证 CRT(conformal radiation therapy) the accuracy of location the accuracy of dose validate
  • 相关文献

参考文献2

共引文献55

同被引文献32

  • 1王鑫,胡超苏,应红梅.放射治疗中的摆位误差[J].肿瘤防治杂志,2004,11(11):1221-1225. 被引量:33
  • 2孔令玲,唐虹.保证三维适形放疗精度的简易方法[J].中华放射肿瘤学杂志,2005,14(2):139-140. 被引量:11
  • 3段诗苗,江浩.鼻咽癌放疗面罩固定技术方法的探讨[J].蚌埠医学院学报,2007,32(2):154-155. 被引量:7
  • 4Mutic S, Dempsey JF, Bosch WR, et al. Multimodality image registration quality assurance for conformal three-dimensional treatment planning[J]. Int Radiat Oncol Boil Phys,2001,51 ( 1 ) : 255 - 260.
  • 5Baroni G,Ferrigno G,Orecchia R,et al.Real time three dimentional motion analysis for patient positioning verification[J].Radiother Oncol,2000,54(1):21-27.
  • 6陈国雄.立体照射与精确放射治疗[A].见:申文江 徐国镇 主编.放射肿瘤学新进展[C].北京:中国医药科技出版社,2001.269-272.
  • 7Lavertu S,Louis MG,Pouliot J,et al.Observation stuty of electronic portal images for offline verification[J].Radiother Oncol,2000,54:47-55.
  • 8Mubata CD,Bidmead AM,Ellingham LM,et al.Portal imaging protocol for radical dose escalated radiotherapy treatment of prostate cancer[J].Int J Radiat Oncol Biol Phys,1998,40:221-231.
  • 9Mohammad Amin MS,Philip ME,William S,et al.A cone beam magavoltage CT scanner for treatment verification in conformal radiotherapy[J].Radiother ncol,1998,48:319-328.
  • 10Guan HQ,Zhu YP.Feasibility of megavoltage portal CT using an electronic portal imaging device (EPID) and a multilevel scheme algebraic reconstruction technique (MLSART)[J].Phys Med Biol,1998,43:2925-2937.

引证文献6

二级引证文献16

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部