摘要
介绍了扫描法测量固体低温比热的实验系统及其出现的问题和解决方法.用此实验系统对纯度为99.7%电介铜的比热测量结果表明,它与美国国家标准参考数据相比,在80K—260K范围内偏差小于7%.
This paper introduced and analysed the experiment system for measuring solid specific heat with the scanning method. Using the experiment system, we measured the specific heat of the copper of 99.7% purity. Its data showed its deviation between 80K and 260K is less than ,l% from the National Standard Reference Data System.
出处
《应用科学学报》
CAS
CSCD
1992年第1期49-52,共4页
Journal of Applied Sciences
关键词
扫描法
比热
测量
固体
微机
specific heat measurement, computer application in experiment.