期刊文献+

1~1000keV能区X射线闪烁探测器灵敏度的标定 被引量:4

CALIBRATION OF X-RAY SCINTILLATION DETECTORS IN 1-KEV TO 1000-KEV ENERGY RANGE
在线阅读 下载PDF
导出
摘要 本文介绍利用六种标准X光同位素源和亚千电子伏X光源,采用直流法对XP1110(或XP1115)光电倍增管与Nal(T1)闪烁体装配成的闪烁探测器在1~1000keV能区范围内的灵敏度进行标定,给出了在三种不同的高压下,不同厚度的新旧两种Nal晶体配制XP1110或XP1115闪烁探测器的标定结果,且给出了这些结果的灵敏度曲线。文章对利用质子激发铝靶国内首次产生的1.5keV的X光作为光源对该闪烁探测器低能道的灵敏度标定,作了简要介绍。文章还对用氮灯作为光源模拟激光打靶对该探测器的幅值线性和积分线性进行了探讨。 In the paper, we introduce the sensitivity calibrations of scintillation detectors made of XP1110(or XP1115)multiplier and Nal(T1) scintillator in the energy range of 1-keV to 1000-keV, utilizing direct - current method and X-ray isotope sources and a subkeV X-ray source. The results of calibrations and the corresponding sensitivity curves of XP1110 or XP1115 scintillators assembled with two kinds of NaI Crystals of different thickness at three different high voltage values are given. The sensitivity calibrations of the scintillator in low-energy channel using a proton - induced fluorescence source are also presented.
出处 《强激光与粒子束》 EI CAS CSCD 1990年第1期23-28,共6页 High Power Laser and Particle Beams
关键词 闪烁探测器 灵敏度 标定 亚千电子伏X光源 幅值线性 积分线性 X射线 sensitivity of scintillator,subkeV X-ray source,amplitude linearity and integral linearity.
  • 相关文献

同被引文献14

  • 1梁齐,高启安,常勇.长方塑料闪烁体光收集效率的研究[J].核技术,1994,17(8):493-498. 被引量:8
  • 2江孝国,王伟,王婉丽.X光作用下CsI:Tl晶体的转换效率研究[J].强激光与粒子束,2005,17(2):203-205. 被引量:13
  • 3GBZ/T144-2002,用于光子外照射放射防护的剂量转换系数[S].
  • 4Jiang H. Computed tomography: Principle, design, artifacts and recent advances[M]. Beijing: Science Press, 2006.
  • 5Chen Ping, Pan Jinxiao, Liu Bin. Industrial computed tomography image size measurement[J].uclear Instruments and Methods in Physics Research Section A, 2009, 608(1) :185-190.
  • 6Constantin F, Pavel C, Bugoi R, et al. An X-ray tomograph based on a flat panel detector[J]. Nuclear Instruments and Methods in Physics Research Section A, 2010, 621(1/3) :685-689.
  • 7Chavanelle J, Parmentier M. A CsI(T1)-PIN photodiode gamma-ray probe[J]. Nuclear Instruments and Methods in Physics Research, 2003, 504(1/3) :321-324.
  • 8复旦大学.核原子物理实验方法[M].北京:原子能出版社,1982:99-105.
  • 9Mark J. Photodetection and measurement., maximizing performance in optical systems[M]. New York. McGraw-Hill, 2003.
  • 10Schlosser D M, Lechner P, Lutz G, et al. Expanding the detection efficiency of silicon drift detectors[J].Nuclear Instruments and Methods in Physics Research Section A, 2010, 624(2) :270-276.

引证文献4

二级引证文献14

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部