期刊文献+

用于宽带取样示波器的Nose-to-nose校准技术 被引量:6

A Calibration Theory Named Nose-to-nose Used on the Broadband Sampling Oscilloscopes
在线阅读 下载PDF
导出
摘要 介绍一种宽带系统校准的新方法 ,对比几种传统的校准方法 ,这种称为nose to nose的校准方法是目前校准精度最高的一种 .这项技术所基于的一个基本事实是 :当示波器的取样电路对直流电压取样时 ,会产生一个极窄的脉冲到输入端 ,这个脉冲正比于系统的冲激响应 ,称为kick out脉冲 ,这样将两台示波器的输入端对接就在第二台示波器上得到了含有两台示波器冲激响应的卷积输出 ,应用反卷积算法可以得到这两台示波器的冲激响应函数 . The nose-to-nose calibration theory is introduced as an accurate method to determine the impulse response of broadband sampling oscilloscopes. The method is based on the fact that when the oscilloscope is sampling a DC-voltage, pulses named kick-out are launched from the sampler towards the input connector and the sampler kick-out is proportional to the oscilloscope′s impulse response. With the method two oscilloscopes input channels are connected together and the second oscilloscope will output the pulses contain information of the two oscilloscopes, applying the deconvolution techniques to the result, the impulse response functions of the two oscilloscopes can be found.
出处 《应用基础与工程科学学报》 EI CSCD 2003年第1期94-97,共4页 Journal of Basic Science and Engineering
基金 北京市教委资助
关键词 宽带取样示波器 Nose-to-nose校准方法 冲激响应函数 反卷积算法 宽带系统校准 取样电路 nose-to-nose, broadband sampling oscilloscope, calibration, impulse response function
  • 相关文献

参考文献6

  • 1刘明亮,高昀,林茂六.“NTN”校准技术中的“Kick-out”脉冲的仿真研究[J].电子测量与仪器学报,2002,16(4):64-68. 被引量:1
  • 2朱江森 刘明亮 高昀.取样电路的非对称性对“NTN”校准技术影响的研究[J].电子测量与仪器学报,2002,16:798-801.
  • 3Verspecht J, Rush K. Individual characterization of broadband sampling oscilloscopes a nose-to-nose calibration procedure[J]. IEEE Trans IM, 1994,43(2) :347-354.
  • 4Riad S M. Modeling of the HP-1430A feed-through wide-band (28-ps) sampling head[J]. IEEE Trans IM, 1982,31 (2):110-115.
  • 5Gans W L. Dynamic calibration of waveform recorders and oscilloscopes using pulse standards [ J ]. IEEE Trans IM, 1990,39 (6) :952-957.
  • 6Verspecht J. Broadband sampling oscilloscope characterization with the " nose-to-nose" calibration procedure: A theoretical and oractice analysis[J]. IEEE Trans IM, 1995,44(6) :991-997.

二级参考文献5

  • 1J. Verspecht and K. Rush. Individual characterization of broad-band sampling oscilloscope with a Nose-to-Nose calibration procedure[J]. IEEE Trans. Instrum. Meas., 1994, Vol. IM- 43(2) :347 - 354.
  • 2J. Verspecht. Broad-band sampling oscilloscope characterization with the "Nose-to-Nose" calibration procedure: a theoretical and practical analysis[J]. IEEE Trans. Instrum. Meas., 1995, Vol. IM-44(6): 991- 997.
  • 3D. F. Williams, K. A. Remley, and D. C. DeGroot. Nose-to-nose response of a 20 GHz sampling circuit[ C]. in 54th ARFTG Conference Digest, 1999, Dec., 1999:64 - 70.
  • 4S.M.Riad. Modeling of the HP-1430A feedthrough wide-band (28ps)sampling head[J]. IEEE Trans. Instrum. Meas.,1982, Vol. IM-31(2): 110- 115.
  • 5刘明亮,林茂六,贺云辉,高剑.基于“NTN”技术的宽带取样示波器的校准[J].计量学报,2002,23(1):65-68. 被引量:14

共引文献1

同被引文献47

引证文献6

二级引证文献17