脉冲调制荧光仪的研制 被引量:2

Design of a Pulse-Amplitude Modulated Fluorometer
在线阅读 下载PDF
导出
摘要 本文在一些通用仪器的基础上自行设计安装了一台脉冲调制荧光仪.在光路处理上用特殊加工的有机玻璃代替光导纤维,有效地克服了以往分叉光导纤维汇合端难以加工得排布均匀致使样品表面激发光照射不均匀的缺点. A Pulse-amplitude modulated fluorometer for photosynthesis research was designed and installed on the basis of several common instruments. In its measurement system, a specially made organic glass column is used in the optical path as a substitute for the optical fiber with three forks, making the distribution of the excitation light on the surface of the leaf more uniform
作者 韩涛 高荣孚
出处 《分析仪器》 CAS 1992年第1期4-8,共5页 Analytical Instrumentation
  • 相关文献

参考文献4

  • 1U. Schreiber,U. Schliwa,W. Bilger. Continuous recording of photochemical and non-photochemical chlorophyll fluorescence quenching with a new type of modulation fluorometer[J] 1986,Photosynthesis Research(1-2):51~62
  • 2W. Bilger,U. Schreiber. Energy-dependent quenching of dark-level chlorophyll fluorescence in intact leaves[J] 1986,Photosynthesis Research(3):303~308
  • 3G. Heinrich Krause,Engelbert Weis. Chlorophyll fluorescence as a tool in plant physiology[J] 1984,Photosynthesis Research(2):139~157
  • 4Ulrich Schreiber. Chlorophyll fluorescence yield changes as a tool in plant physiology I. The measuring system[J] 1983,Photosynthesis Research(1):361~373