高纯氧化镝电感耦合等离子体原子发射光谱法摄谱分析时光谱干扰校正方法的研究 被引量:8

Correction of Spectral Interference in the Analysis of High Purity Dysprosia by Inductively Coupled Plasma-Spectrography
在线阅读 下载PDF
导出
摘要 本文建立了高纯氧化镝样品ICP摄谱分析时干扰系数法校正光谱干扰的方法。测定了镝基体对14种稀土杂质元素20条灵敏分析线的干扰系数(KCT)。并有效地校正了镝基体造成的光谱干扰,获得满意的结果。 The present paper deals with the development of the interference coefficient method for correcting the spectral interferences in the analysis of high purity Dy_2O_3 with ICP-AES spectrography. The interference coefficients of twenty sensitive analysis lines have been determined. The spectral interferences from Dy matrix have been effectively corrected. Using the correction method, the recovery of the impurities in high purity Dy_2O_3 is 85%~115%.
机构地区 南开大学化学系
出处 《分析化学》 SCIE EI CAS CSCD 北大核心 1992年第5期600-603,共4页 Chinese Journal of Analytical Chemistry
基金 国家自然科学基金
关键词 光谱干扰 稀土 ICP-AES Inductively coupled plasma, Spectral interference, Atomic emission spectrometry, Rare earth element.
  • 相关文献

参考文献3

  • 1Zhang Ronglin,1990年
  • 2黄志荣,理化检验.化学分册,1988年,24卷,6期,325页
  • 3陈新坤,电感耦合等离子体光谱法原理和应用,1987年