摘要
本文将流动注射分析(FIA)引入Mo-SCN^--RhB三元缔合体系测定痕量Mo。不需萃取,反应迅速,重现性好。0~1.6μg/ml Mo符合比耳定律,检测限0.02μg/ml。用预混线圈R_1可克服Fe^(3+)等元素与SCN^-络合的干扰。测定了植物中Mo,结果满意。
A method of flow injection analysis (FIA) was introduced into the spectrophotometric system of Mo-SCN^--RhB ternary complex for the determination of trace Mo. This method does not need preliminary extraction, and was a rapid, reproducible one, Beer's law was obeyed within the range from 0.00 to 1.60 μg/ml for Mo, and the detection Iimit was 0.02 μg/ml. The loop R_1 was used so that the interference of iron complex could be eliminated. The determination of Mo in plants was carried out with satisfactory results.
出处
《分析化学》
SCIE
EI
CAS
CSCD
北大核心
1992年第3期319-321,共3页
Chinese Journal of Analytical Chemistry
关键词
流动注射分析
分光光度法
钼
植物
Flow injection analysis, Spectrophotometry, Molybdenum, Plants.