期刊文献+

流动电解池-介质交换法分离重叠溶出峰的研究——痕量铅、锡的同时测定 被引量:2

Separation of Overlapping Stripping Peaks With a Flow-through Cell and Medium Exchange Method——Simultaneous Determination of Trace Tin and Lead
在线阅读 下载PDF
导出
摘要 本文采用流通电解池一介质交换方法研究了铅和锡在酸性介质中的差示脉冲阳极溶出伏安行为,在流动的含有样品的盐酸溶液中预电解富集,而在静止的不含样品的草酸溶液中溶出,其△Ep≥90 mV,波形完好。确定了同时测定痕量级铅、锡的最佳条件,样品分析结果表明,方法准确、简便、快速、灵敏。 Differential pulse anodic stripping voltammetry with a flow-through cell and medium exchange is studied to separate the oxidation peaks of tin and lead in acidic medium, which involves pre-electrolysis in hydrochloric acid medium and stripping in oxalic acid medium. Both peaks of tin and lead have been separated well with the difference of 90 mV. Experimental results show the dependence of the peak current on the pre-electrolysis time, pre-electrolysis potential, solution flow rate and conditions of stripping medium. This method has been applied to the determination of tin and lead in natural weter samples with satisfactory results.
出处 《分析化学》 SCIE EI CAS CSCD 北大核心 1992年第3期352-355,共4页 Chinese Journal of Analytical Chemistry
关键词 流通电解池 介质交换 差示脉冲 Flow-through cell, Medium exchange, Differential pulse, Tin, Lead.