期刊文献+

输血输液加温设备不良事件的分析 被引量:2

Analysis of Adverse Events of Blood Transfusion and Infusion Heating Equipment at Home and Abroad
暂未订购
导出
摘要 输血输液加温器是临床用于加热血液、药液等输注液体的关键设备,加热方式主要包括干热式和循环水式。该研究通过分析国内外文献及不良事件监测数据,发现输血输液加温器主要故障包括加温异常、温度偏差、电子元件故障及液体泄漏等。风险原因涉及设计缺陷、操作不当、维护不足及质量控制等问题。建议采取优化设计、提升温度精度、加强维护培训及完善质控体系等措施,以提高设备安全性和可靠性。该研究可为国家“十四五”医疗器械不良事件重点监测提供参考,助力临床用械安全。 The purpose of this study was to analyze the characteristics and related factors of adverse events of blood transfusion infusion heaters,and to provide reference for the use of clinical medical devices.Through systematic statistical analysis of relevant literature at home and abroad,and retrieval of medical device alert database,the causes of adverse events are sorted out,and then the potential risks of blood transfusion heaters are analyzed,and corresponding improvement measures are proposed.It is the key measure to improve the safety of equipment,standardize the use rationality of clinical medical devices,and prevent the occurrence of serious adverse events.
作者 宗冉 张莹 牛惠芳 赵玉娟 宋来全 刘金花 Zong Ran;Zhang Ying;Niu Huifang;Zhao Yujuan;Song Laiquan;Liu Jinhua(Ji'ning Institute for Food and Control,Shandong Jining 272000,China;Shandong Center for ADR Monitoring,Shandong Jinan 250000,China;Ji'ning First People's Hospital,Shandong Jining 272000,China)
出处 《医疗装备》 2025年第6期34-36,共3页 Medical Equipment
关键词 输血输液加温器 不良事件 故障 Blood transfusion infusion warmer Adverse events Fault
  • 相关文献

参考文献12

二级参考文献106

共引文献70

同被引文献13

引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部