期刊文献+

紫外规则透射比国际比对 被引量:1

The Comparison of Ultraviolet Regular Transmittance
在线阅读 下载PDF
导出
摘要 为保证我国紫外透射量值与国际的一致性,中国计量科学研究院代表我国参加了在2018年至2019年间,欧亚计量合作组织(COOMET)组织的首次紫外透射比的研究性比对。这次比对采用了4组溶液作为比对样品,透射比量值标称值覆盖5%~90%,波段覆盖235~350 nm。比对结果表明,各参比实验室测量结果基本一致。在透射比50%附近,我国与距离国际参考值的偏离量最大不超过0.12%;在透射比20%附近,我国与距离国际参考值的偏离量不足0.08%;在透射比10%附近,我国与距离国际参考值的偏离量不超过0.03%。 In order to ensure the consistency of UV transmittance values in China with those in the world,the National Institute of Metrology,China,participated in the first research comparison of ultraviolet regular transmittance organized by COOMET during 2018 to 2019.This comparison uses 4 sets of solutions as comparison samples,the nominal value of transmittance covers 5%~90%,and the wavelength range covers 235nm~350nm.The comparison results show that the measurement results of each reference laboratory are basically equivalent and consistent.When the transmittance is 50%,the deviation between ours and the international reference value can not exceed 0.12%.When the transmittance is near 20%,the deviation between ours and the international reference value is less than 0.08%.For 10%transmittance,the deviates quantity is no more than 0.03%.
作者 冯国进 张巧香 郑春弟 甘海勇 吴厚平 李雨宵 梁凤臣 FENG Guojin;ZHANG Qiaoxiang;ZHENG Chundi;GAN Haiyong;WU Houping;LI Yuxiao;LIANG Fengchen(National Institute of Metrology,Beijing 100029,China)
出处 《计量学报》 CSCD 北大核心 2024年第8期1103-1107,共5页 Acta Metrologica Sinica
基金 国家重点研发专项(2021YFF0600204)。
关键词 光学计量 紫外透射比 不确定度 重铬酸钾溶液 国际比对 optical metrology ultraviolet transmittance uncertainty potassium dichromate solution international comparison
  • 相关文献

参考文献4

二级参考文献16

共引文献3

同被引文献12

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部