期刊文献+

低压离子色谱法测定三苯基膦中痕量氯离子 被引量:2

Determination of Trace Amounts of Chloride ion in Triphenyl Phosphine by Low-Pressure Ion Chromatography
在线阅读 下载PDF
导出
摘要 采用双柱低压离子色谱分离,电导检测了三苯基膦样品中的痕量氯离子.以Na2CO3作淋洗液,流动注射进样,优化色谱分离条件,并测定了该法的精密度,线性范围及回收率. A method was developed for the determination of trace amounts of chloride ion in triphenyl phosphine sample by lowpressure ion chromatography with conductive detection. The eluent used was 15 mmol/L NaCO3. Calibration curves showed good linearity with r=0999 4(at 001-4 mg/L). The detection limit of chloride by this method was 5 g/L. This method is simple and accurate, in good stability and sensitivity, so it has been used satisfactorily to analyze triphenyl phosphine sample in factory.
出处 《西南师范大学学报(自然科学版)》 CAS CSCD 北大核心 2002年第6期910-912,共3页 Journal of Southwest China Normal University(Natural Science Edition)
基金 99'教育部春晖计划资助项目.
关键词 低压离子色谱法 氯离子 三苯基膦 催化剂 含量测定 石油化工 chloride ion triphenyl phosphine low-pressure ion chromatography