基于临床使用工况的磁共振(MR)可靠性指标与实现研究

Research on Reliability Index and Realization of Magnetic Resonance(MR)Based on Clinical Use Condition
暂未订购
导出
摘要 该研究结合上海市第六人民医院在用MR的临床使用工况,将MR部件分为扫描Ⅰ型和扫描Ⅱ型。同时,结合MR部件的主失效应力,将MR的分为动力学部件和电热学部件。该研究给出了系统可靠性指标确定、系统可靠性分配、部件可靠性指标实现、系统可靠性预计和系统可靠性验证一整套的磁共振(以下简称"MR")系统可靠性指标与实现方法。同时,该研究也给出了可靠性预计和可靠性验证的方法,并给出了基于临床使用数据统计的MR系统MTBF计算方法。 Combined with the clinical use condition of MR in use in Shanghai Sixth People’s Hospital, MR components are divided into scanning type Ⅰ and scanning type Ⅱ. At the same time, combined with the main loss force of MR components, the research divides MR components into dynamic components and electric thermal components. In this study, a complete set of MR system reliability indexes and implementation methods are given, including system reliability index determination, system reliability allocation, component reliability index realization, system reliability prediction and system reliability verification. At the same time, this study also gives the methods of reliability prediction and reliability verification, and gives the MTBF calculation method of MR system based on clinical use data statistics.
作者 贾东方 贺飞 金玮 张尉强 刘义红 邹金林 JIA Dongfang;HE Fei;JIN Wei;ZHANG Weiqiang;LIU Yihong;ZOU Jinlin(Shanghai United Imaging Healthcare Co.,Ltd,Shanghai,201800;Shanghai Institute of Medical Device Testing,Shanghai,201318;The Sixth People’s Hospital Affiliated to Shanghai Jiao Tong University,Shanghai,200233;Re-impro Reliability Technology Co.Ltd.,Guangzhou,510610)
出处 《中国医疗器械杂志》 2021年第6期628-635,共8页 Chinese Journal of Medical Instrumentation
基金 医学影像设备可靠性与工程化技术研究及应用项目(2018YFC0115800)
关键词 磁共振 临床使用工况 寿命加速试验 加速因子(AF) magnetic resonance imaging clinical use condition accelerated life test acceleration factor(AF)
  • 相关文献

参考文献6

二级参考文献46

共引文献56

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部