摘要 作为核心检测器件,光电倍增管(PMT)被广泛用于化学发光免疫分析仪中,具有噪声低、灵敏度高、线性范围宽的特点,其是否能够正常工作决定了检测结果的准确性。该研究根据化学发光免疫分析仪的原理和技术特性,设计了一款PMT校验装置,用于测试PMT功能是否正常,该校验装置不影响PMT作为探测器件正常的检测功能,线性好,可有效检测PMT的性能;此外,其可在化学发光免疫分析仪每次使用前对PMT进行线性度检测,评估PMT功能是否正常,并对PMT失效进行警示,有效提高了仪器的可靠性。 Photomultiplier tubes(PMT)are widely used in chemiluminescence immunoanalyzer,with the characteristics of low noise sensitivity and high sensitivity and wide linear range.As a core detection device,the normal operation of the device determines the accuracy of the detection results.According to the principle and technical characteristics of chemiluminescence immunoanalyzer,a PMT calibration device was designed in this study to test whether the PMT function is normal.It does not affect the normal detection function of PMT as a detector.The device has good linearity and can effectively detect the performance of PMT.Besides,the device can linearly detect the PMT before the chemiluminescence immunoanalyzer was used every time to evaluate whether the PMT function is normal and alarm the device failure,effectively improving the reliability of the instrument.
作者 邓振进 周阳 刘向荣 邬鹏程 高飞 Deng Zhenjin;Zhou Yang;Liu Xiangrong;Wu Pengcheng;Gao Fei(Hunan Testing Institute For Medical Devices,Changsha Hunan 410011,China;Hunan Legendic Biotechnology Co.,Ltd,Changsha Hunan 410205,China)
出处 《医疗装备》 2021年第21期4-5,共2页 Medical Equipment
基金 创新研究群体项目科药联合基金项目(2020JJ9040)。
关键词 光电倍增管 光电探测器件 参考光源 Photomultiplier Photoelectric detector Reference light source
  • 相关文献

参考文献1

二级参考文献22

  • 1Shaw G A, Siegel A M, Model J. Ultraviolet comm links for distributed sensor networks [ C ]//Digest of the LEOS Summer Topical Meetings,2005 : 39 - 40.
  • 2XU Zhengyuan, BRIAN M Sadler. Ultraviolet communica- tions : potential and state of the art [ C ]//IEEE Communi- cation Magazine ,2008,46 ( 5 ) :67 - 73.
  • 3冯兴平.紫外光通信信道散射模型研究及实验系统设计[D].成都:电子科技大学,2009.
  • 4Lindner,Malka. Solar blind bandpass filters for UV ima- ging devices [ J ]. SPIE, 1998, (3302) : 176 - 183.
  • 5R G Satin, I S Ga nutdinov, R S Sabirov, et al. Solar-blind filter for the ultraviolet region [ J ]. J. Opt. Technol, 2007, 74(3) :208 -210.
  • 6Sangsik Kim, Mengren Man, Minghao Qi, et al. Multilayer metal-dielectric stack ultraviolet filter[ C ]. Laser Science, 2013.
  • 7Sangsik Kim, Mengren Man, Minghao Qi, et al. Angle-in- sensitive and solar-blind ultraviolet and pass filter [ J ]. Optics Letters ,2014,39(19) :5784 - 5787.
  • 8闫浩.日盲型紫外探测系统滤光膜的研制[D].长春:长春理工大学.2012.
  • 9付联效,吴永刚,郑秀萍,王振华,彭东功,焦宏飞,曹鸿,吕刚,周洪军,霍同林.中心波长121.6nm的真空紫外窄带滤光片设计和制备[J].光学学报,2008,28(12):2427-2430. 被引量:7
  • 10许智煌,庄欣欣,苏根博.有机紫外滤波材料的紫外光谱性质研究[J].光学学报,2009,29(7):1918-1920. 被引量:7