体位改变时经颈内静脉植入胸壁式输液港导管头端位置变化及影响因素 被引量:5

Movement of catheter tip of chest wall access port implanted via internal jugular vein and impact factors when changing body posture
暂未订购
导出
摘要 目的观察由卧位至立位经颈内静脉植入胸壁式输液港导管头端位置变化,并分析其影响因素。方法回顾性分析171例接受颈内静脉植入胸壁港患者,均于术后拍摄卧位和立位X线胸片,测量由卧位至立位时导管头端移动距离,采用Pearson相关分析及多元回归分析观察导管头端移动距离的影响因素。结果由卧位至立位时导管头端呈向头侧移位倾向,平均移动距离(15.82±16.87)mm。导管头端移动距离与性别(P=0.039)及皮脂厚度(P=0.011)呈正相关,与留置导管长度(P=0.039)呈负相关。结论由卧位至立位时,经颈内静脉植入胸壁式输液港导管头端倾向于向头侧移位,移动距离与患者性别、皮脂厚度及留置导管长度相关。 Objective To observe movement of catheter tip of chest wall access port implanted via the internal jugular vein and impact factors when changing from supine to vertical positions.Methods Data of 171 patients underwent implantation of chest wall access port via the internal jugular vein and chest X-ray films on supine and vertical position postoperatively were retrospectively analyzed,and the distance from the tracheal protuberance to catheter tip was measured.Pearson correlation analysis and multivariate regression analysis were performed to observe the impact factors of moving distance of catheter tip.Results Changing from supine to vertical position,the catheter tip was inclined to cephalad movement,and the mean moving distance was(15.82±16.87)mm.The distance of catheter tip movement was positively correlated with gender(P=0.039)and sebum thickness(P=0.011)and negatively correlated with indwelling length of catheter(P=0.039).Conclusion The catheter tip of chest wall access port implanted via the internal jugular vein tends to shift cephalad when changing from supine to vertical position,and the distance of movement is related to patient’s gender,sebaceous thickness and the length of indwelling catheter.
作者 俞靖凡 金泳海 樊宝瑞 杜鹏 李明明 张申 倪才方 YU Jingfan;JIN Yonghai;FAN Baorui;DU Peng;LI Mingming;ZHANG Shen;NI Caifang(Department of Vascular and Interventional Radiology,the First Affiliated Hospital of Soochow University,Suzhou 215006,China)
出处 《中国介入影像与治疗学》 北大核心 2020年第7期411-414,共4页 Chinese Journal of Interventional Imaging and Therapy
关键词 植入式静脉输液港 血管造影术 数字减影 体位 implantable venous access port angiography,digital subtraction posture
  • 相关文献

参考文献5

二级参考文献66

  • 1余秀华,施红,许韵韵,李黎.B超测量腹部脂肪的研究[J].中华超声影像学杂志,1994,3(2):67-69. 被引量:6
  • 2谢嘉平,任力锋,黄穗红,刘喜玲.振动减肥中腹壁脂肪的超声检测与力学特征[J].中国医学物理学杂志,1996,13(4):213-216. 被引量:4
  • 3许秀芳,李晓蓉,刘玉金.肿瘤介入护理学[M].北京:科学出版社,2011:32.
  • 4程永德,程英升,颜志平,等.常见恶性肿瘤介入治疗指南[M].北京:科学出版社,2013年.
  • 5Niederhuber JE, Ensminger W, Gyves JW, et al. Totally implanted venous and arterial access system to replace external catheters in cancer treatment[J]. Surgery, 1982, 92: 706-712.
  • 6Kreis H, Loehberg CR, Lux MP, et al. Patients' attitudes to totally implantable venous access port systems for gynecological or breast malignancies[J]. Eur J Surg Oncol, 2007, 33: 39-43.
  • 7Ignatov A, Hoffman O, Smith B, et al. An 11-year retrospective study of totally implanted central venous access ports: com- plications and patient satisfaction [J]. Eur J Surg Oncol, 2009, 35:241-246.
  • 8Ruesch S, Walder B, Tramer MR. Complications of central venous catheters: internal jugular versus subclavian access: a systematic review[J]. Crit Care Med, 2002, 30: 454-460.
  • 9Lorente L, Henry C, Martin MM, et al. Central venous catheter- related infection in a prospective and observational study of 2,595 catheters[J]. Crit Care, 2005, 9: R631-R635.
  • 10American Society of Anesthesiologists Task Force on Central Venous Access, Rupp SM, Apfelbaum JL, et al. Practice guidelines for central venous access: a report by the American Society of Anesthesiologists Task Force on Central Venous Access [ J]. Anesthesiology, 2012, 116: 559-573.

共引文献184

同被引文献39

引证文献5

二级引证文献20

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部