期刊文献+

现代麻醉机检测方法 被引量:2

Detection Method of Modern Anesthesia Machine
在线阅读 下载PDF
导出
摘要 填补麻醉机检测空缺,对麻醉机及麻醉工作站进行全面地质量控制检测,确保其安全可靠,降低医疗事故的风险,以《YY 0635.4-2009吸入式麻醉系统第4部分:麻醉呼吸机》等标准为依据,对潮气量、气道峰压、呼气末正压、呼吸频率、吸气氧浓度、气体流量计流速、麻醉气体浓度和脉搏-血氧饱和度等参数进行检测,完成了麻醉机及麻醉工作站的检测,推进医疗设备的有效质量控制。 In order to fill in the vacancy of anesthesia machine testing, and carry out comprehensive quality control testing on anesthesia machine and anesthesia workstation, so as to ensure its safety and reliability and reduce the risk of medical accidents. Based on the standards of YY 0635.4-2009 inhalation anesthesia system Part 4: anesthesia ventilator, the parameters of tidal vol- ume, peak airway pressure, positive end expiratory pressure, respiratory frequency, inspiratory oxygen concentration, gas flowmeter flow rate, anesthetic gas concentration and pulse blood oxygen saturation were measured. The detection of anesthesia machine and anesthesia workstation was completed to promote the effective quality control of medical equipment.
作者 唐浩然 Tang Haoran(Guangzhou Radio and Television Measurement and Testing Co.,Ltd.,Guangzhou,510000,China)
出处 《仪器仪表用户》 2020年第7期102-104,共3页 Instrumentation
关键词 麻醉机 质量控制 检测 anesthesia machine quality control detection
  • 相关文献

参考文献4

二级参考文献22

共引文献24

同被引文献13

引证文献2

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部