摘要
填补麻醉机检测空缺,对麻醉机及麻醉工作站进行全面地质量控制检测,确保其安全可靠,降低医疗事故的风险,以《YY 0635.4-2009吸入式麻醉系统第4部分:麻醉呼吸机》等标准为依据,对潮气量、气道峰压、呼气末正压、呼吸频率、吸气氧浓度、气体流量计流速、麻醉气体浓度和脉搏-血氧饱和度等参数进行检测,完成了麻醉机及麻醉工作站的检测,推进医疗设备的有效质量控制。
In order to fill in the vacancy of anesthesia machine testing, and carry out comprehensive quality control testing on anesthesia machine and anesthesia workstation, so as to ensure its safety and reliability and reduce the risk of medical accidents. Based on the standards of YY 0635.4-2009 inhalation anesthesia system Part 4: anesthesia ventilator, the parameters of tidal vol- ume, peak airway pressure, positive end expiratory pressure, respiratory frequency, inspiratory oxygen concentration, gas flowmeter flow rate, anesthetic gas concentration and pulse blood oxygen saturation were measured. The detection of anesthesia machine and anesthesia workstation was completed to promote the effective quality control of medical equipment.