医用X射线诊断设备生产企业调试机房辐射水平测量及分析 被引量:1

Measurement and analysis of radiation level in debugging machine room of the manufacturing enterprise of medical X-ray diagnostic equipment
暂未订购
导出
摘要 目的:对医用X射线诊断设备生产企业的调试机房周围辐射水平进行检测,分析个别点位剂量率水平较高的原因,制定解决方案,以提高调试机房的辐射防护水平。方法:参照国家标准《医用X射线诊断放射防护要求》对北京地区具有代表性的医用X射线诊疗设备生产企业的医用X射线诊断设备调试机房周围辐射水平进行检测,对调试工程师的个人剂量进行监测,掌握调试机房外辐射水平和调试工程师的辐射防护情况。结果:检测的企业医用X射线诊断设备调试机房外辐射水平和调试工程师的个人剂量监测结果均符合国家标准的要求。结论:加强医用X射线诊断设备调试机房辐射防护的企业管理和政府监管,可有效降低调试机房周围辐射水平,并为做好调试工程师的辐射防护提供参考。 Objective:To implement test for radiation level around the debugging room of manufacturing enterprise of medical X-ray diagnostic equipment so as to improve the protection level for radiation around debugging machine room.Methods:According to national standard<Requirements for medical X-ray diagnosis and radiation protection>,the radiation level around the debugging machine room of medical X-ray diagnosis equipment of manufacturing enterprise with representativeness of medical X-ray diagnosis and treatment equipment in Beijing region was tested,and the personal dose of the debugging engineer was monitored.The radiation levels outside the debugging machine room were grasped,and the radiation protection situation of the debugging engineer was monitored.Results:Both the radiation level outside the debugging machine room of medical X-ray diagnostic equipment of tested enterprise and the monitoring results of personal dose of debugging engineer met the requirements of national standard.Conclusion:It is very important to strengthen the enterprise management and government supervision for radiation protection of debugging machine room of medical X-ray diagnostic equipment,and it can effectively reduce the radiation level around debugging machine room and provide reference for achieving well radiation protection of debugging engineers.
作者 王志超 郭文 张贵英 丁艳秋 郝述霞 WANG Zhi-chao;GUO Wen;ZHANG Gui-ying(Beijing Prevention and Treatment Hospital of Occupational Disease for Chemical Industry,Beijing 100093,China)
出处 《中国医学装备》 2020年第2期9-12,共4页 China Medical Equipment
基金 科技部科研院所技术开发研究专项(2012EG150137)“多功能被动式个人剂量全自动测量系统的研制”
关键词 医用X射线 诊断设备 调试机房 辐射水平 生产企业 Medical X-ray Diagnostic equipment Debugging room Radiation level Manufacturer
  • 相关文献

参考文献5

二级参考文献27

  • 1罗明泉 郑桂芳.辐射危害与防护[M].西安:陕西科学技术出版社,1998.118-119.
  • 2GB18871-2002.电离辐射防护与辐射源安全基本标准[S].[S].,..
  • 3GBZ130-2002.医用X射线诊断卫生防护标准[S].[S].,..
  • 4GBZ138-2002.医用X射线诊断卫生防护监测规范[S].[S].,..
  • 5Michael. Commissioning of a Mobile Electron Acceleratorfor Intraoperative Radiotherapy. J Applied Clin Med Phy, 2001,2(3) :121 - 130.
  • 6Beddar AS, Biggs P J, Chang S, et al. Intraoperative radiation therapy using mobile electron linear accelerators. Report of AAPM Radiation Therapy Committee Task Group No. 72. Med Phys,2006, 33 : 1476 - 1489.
  • 7Alexandre S, Krechetov, Donald Goer. Shielding assessment of a mobile electron accelerator for intraoperative radiotherapy. Journal of Applied Clinical Medical Physics, 2010,11 ( 4 ) : 1 - 11.
  • 8A. Sam Beddar, Sunil Krishnan. Intraoperative radiotherapy using a mobile electron LINAC: A retroperitoneal sarcoma case. J Applied Clin Med Phy, 2005,6(3) :95 - 106.
  • 9Structural Shielding Design and Evaluation for Megavoltage X- and Gamma-Ray Radiotherapy facilities. NCRP report 151, 2007.
  • 10Dares JL, Mills MD. Shielding assessment of a mobile electron accelerator for intraoperative radiotherapy. J Applied Clin Med Phy,2001, 2(3) : 165 - 173.

共引文献27

同被引文献17

引证文献1

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部