骨密度量值溯源系统设计研究 被引量:3

Research on the Design of Bone Mineral Density Traceability System
在线阅读 下载PDF
导出
摘要 双能X射线骨密度仪(Dual Energy X-ray Absorptiometry,DEXA)测量获得的骨密度值是用于骨质疏松症诊断的国际金标准,但不同品牌DEXA测量结果的一致性和骨密度值的准确性及溯源方法尚未获得足够关注。为统一DEXA的测量结果,同时为骨质疏松症临床诊断结果互认提供技术支持,本文拟设计并搭建一套多能量X射线骨密度模体校准装置,研究DEXA骨密度模体计量检测和溯源方法,解决DEXA测量骨密度的量值统一和溯源问题。目前,该校准装置已完成设计与加工,经试验验证其辐射剂量等稳定性指标优于1%。 Bone mineral density measured by dual energy X-ray absorptiometry(DEXA)is international Gold Standard used for diagnosing osteoporosis.However,the accuracy and traceability of bone mineral density measured by different DEXA from different manufactures did not gain enough attention.To unify the measurement results from different DEXA and supply technique support to clinical diagnostic results of osteoporosis,a set of calibration equipment for DEXA was designed to realize the metrological test and tracing method of bone mineral density phantom as well as the quantitative uniformity and traceability of bone mineral density measured by DEXA.The calibration equipment has been designed and manufactured,and its stability of radiation dosage is proved to be better than 1%by experiments.
作者 高明亮 李成伟 万国庆 刘文丽 GAO Mingliang;LI Chengwei;WAN Guoqing;LIU Wenli(Division of Medical and Biological Metrology,National Institute of Metrology,Beijing 100029,China)
出处 《中国医疗设备》 2019年第11期7-10,29,共5页 China Medical Devices
基金 国家重点研发计划专项(2016YFF0201003)
关键词 双能X射线骨密度仪 骨密度值 多能量X射线骨密度模体校准装置 溯源 dual energy X-ray absorptiometry bone mineral density multi-energy X-ray calibration equipment for standard BMD phantom traceability
  • 相关文献

参考文献9

二级参考文献93

共引文献103

同被引文献50

引证文献3

二级引证文献26

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部