摘要
目的 :探讨XG5 0 1型遥控X线机故障原因和检查方法。方法 :应用排除的方法检查电路 ,对移相电路工作原理作理论性分析。结果 :理论分析的结果与实际测试相符。结论 :采用排除法进行理论分析 ,并用示波器测试电路 。
Objective To evaluate the cause of trouble and the examination method in XG 501 Remote Contral X-ray Machine. Methods The circuit was examined with a method of exclusion, and the work principle of phaser circuit was theoretically analysed. Results The results of theoretical analysis answer to actual measure. Conclusion Theoretical analyses by a method of exclusion, and measure the circuit with oscillograph have important meaning for the trouble was correctly judged and the machine was reinstated.
出处
《医疗装备》
2002年第9期46-47,共2页
Medical Equipment