期刊文献+

硅酸根分析仪示值误差测量结果的不确定度评定

Evaluation of Uncertainty in Measurement Results Error of Silicate Analyzer
在线阅读 下载PDF
导出
摘要 文中为了规范硅酸根分析仪的校准过程,保证溯源有效性,以示值误差建立不确定度评定模型,计算校准过程的不确定度。依据硅酸根分析仪校准规范,从测量重复性、分辨力、二氧化硅标准溶液的不确定度、稀释过程使用的玻璃量具等角度出发,分析校准过程中引入的各类不确定度,计算其灵敏度,分别评定>100μg/L及≤100μg/L量程的硅酸根分析仪校准结果不确定度,确保其在环境保护、安全生产方面提供数据保证。 In order to standardize the calibration process of thesilicate analyzer and ensure the validity of the traceability, this paper establishes the uncertainty evaluation model of the value error and calculates the uncertainty of the calibration process. Based on the silicate analyzer calibration specifications, from the perspective of measuring repeatability and resolution, the uncertainty of the silicon dioxide standard solution, and the measuring glass in the process of dilution, the author analyzes the uncertainty of calibration process, calculates its sensitivity and assesses the uncertainty of calibration results of silicate analyzer, which the range is greater than 100μg/Land less than or equal to100μg/L, to ensure that the data provides the basis in terms of environmental protection and safety in production.
出处 《质量技术监督研究》 2017年第5期2-5,共4页 Quality and Technical Supervision Research
关键词 硅酸根分析仪 不确定度 示值误差 Silicate analyzer Uncertainty Measurement results error
  • 相关文献

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部