射波刀颅骨追踪俯卧位与仰卧位治疗精度对比 被引量:4

Skull tracking accuracy of prone and supine positions in Cyberknife treatment: a comparison study
暂未订购
导出
摘要 目的:探讨射波刀颅骨追踪俯卧位与仰卧位治疗精度的差异,为射波刀颅骨追踪俯卧位治疗提供理论依据。方法:分别对安装有EBT3免冲洗胶片的头颈模体采用俯卧位与仰卧位进行CT扫描、计划设计和模拟射波刀治疗。采用E2E软件分析曝光后的EBT3免洗胶片,对比射波刀颅骨追踪俯卧位与仰卧位治疗精度。结果:头颈模体俯卧位与仰卧位的治疗精度分别为0.81和0.72 mm,均满足ACCURAY公司提供的0.90 mm的治疗标准,两者治疗精度仅差0.09 mm,可认为头颈模体颅骨追踪俯卧位与仰卧位具有同等的照射精度。结论:当患者无法进行仰卧位颅骨追踪射波刀治疗时,可以采用俯卧位颅骨追踪方式进行治疗。 Objective To compare the differences in skull tracking accuracy between prone and supine positions in Cyberknife treatment, and to provide the theory basis for Cyberknife treatment in prone position for skull tracking. Methods CT scans were performed, and the treatment plans were designed and executed for the head and neck phantom with EBT3 film in prone and supine positions, respectively. The EBT3 films after exposure were analyzed with E2 E software to compare the Cyberknife treatment accuracy between prone and supine positions for skull tracking. Results The treatment accuracies for head and neck phantom in prone and supine positions were 0.81 and 0.72 mm, respectively, better than the treatment standard provided by ACCURAY company(0.90 mm). The treatment accuracy difference between prone and supine positions was only 0.09 mm,therefore, the treatment accuracies of prone and supine positions were considered equal. Conclusion When patients cannot receive Cyberknife treatment for skull tracking in supine position, Cyberknife treatment can be applied for patients in prone position.
出处 《中国医学物理学杂志》 CSCD 2017年第8期790-793,共4页 Chinese Journal of Medical Physics
关键词 射波刀 颅骨追踪 俯卧位 仰卧位 治疗精度 Cyberknife skull tracking prone position supine position treatment accuracy
  • 相关文献

参考文献5

二级参考文献34

  • 1国家质量监督检验检疫总局.GB18871-2002电离辐射防护与辐射源安全基本标准[S].北京:中国标准出版社,2003.
  • 2李玉,韩悦荣,刘骁雄.射波刀//李玉,梁军.三维适形与词强放射治疗的基础与临床[M].北京:科学技术出版社,2010:291-295.
  • 3梁军,李玉,柴广金.图像引导技术与肝胆胰肿瘤放射治疗//李玉.肝胆胰恶性肿瘤的微创新技术[M].北京:科学技术出版社,2009:86-97.
  • 4West JB,Fitzpatrick JM,Toms SA,et al. Fiducial point placement and the accuracy of point-based,rigid body registration[J]. Neurosurgery,2001, 48:810-817.
  • 5Khan FM. The physics of radiation therapy,lippincott,williams and wilkins, baltimore ,MD[J].Quality Assurance ,2003,17:315-317.
  • 6Hurkmans CW, Remeijer P, Lebesque JV ,et al. Set-up verification using portal imaging:review of current clinical practice,radiother [J].Oncol, 2001,58 : 205 -216.
  • 7Sonja D. Quality assurance for robotic radiosurgery :report of the AAPM task group 135[J]. Medical Physics, 2011,38 (6) : 2924-2926.
  • 8Quinn AM. CyberKnife: a robotic radiosurgery system[ J ]. Clin J OncolNurs, 2002,6 : 149 - 156.
  • 9Kothary N, Heit J J, Louie JD, et al. Safety and efficacy of percuta- neous fiducial marker implantation for image - guide radiation [ J ]. J Vasc Interv Radiol, 2009,20( 2 ) :235 - 239.
  • 10Sotiropoulou E, Stathochfistopoulou I, Stathopoulos K, et al. CT- guided fiducial placement for cyberknife stereotactic radiosurgery: an initial experience [ J ]. Cardiovasc Intervent Radiol, 2010,33 : 586 - 589.

共引文献39

同被引文献29

引证文献4

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部