摘要
为准确、有效地解决固定式剂量仪器的量值溯源,构建基于G函数法的环境级别的剂量装置模型。通过使用MC法,对一款Na I(Tl)探测器的脉冲幅度分布谱进行模拟,并据此考察了截断能量Emin、阶数K、道宽ΔE三个因素对G函数及其计算结果的影响,发现截断能量Emin、阶数K、道宽ΔE对G函数的形状及剂量率计算结果均会产生影响。
In order to accurately and efficiently solve the traceability of the stationary dose rate instruments, a Monte Carlo model of measuring dose rate used in environment level based on G - function method was buih. Then pulse -height distribution spectrum detected by a NaI (T1) detector was simulated by using the Monte Carlo method. And the results show that E,.in - the energy limited, K - the order of G - function and AE - the channel width of MCA could affect the accuracy of the dose rate got from using G - function in different way. And the final determination of the three parameters should be depended on experimental condition.
出处
《核电子学与探测技术》
CAS
北大核心
2016年第3期283-286,共4页
Nuclear Electronics & Detection Technology
关键词
G函数
MC方法
剂量(率)
G - function, Monte Carlo Method, dose( - rate)