Establishment of Reference Intervals and Transfusion Criterion for Sonoclot Analysis 被引量:4

Establishment of Reference Intervals and Transfusion Criterion for Sonoclot Analysis
暂未订购
导出
摘要 Sonoclot analyzer has been widely used in many countries. But the reference intervals provided by the manufacturer were derived from only 45 participants, and there was no cut-off value for transfusion for Sonoclot analysis. This study aimed to establish reference intervals and transfusion criterion for Sonoclot analysis. Volunteers were recruited from healthy Chinese adults and patients undergoing cardiac surgery. Blood samples were withdrawn from forearm vein and measured for activated clotting time(ACT), clot rate(CR), platelet function(PF), activated partial thromboplastin time(APTT), fibrinogen concentration(FIB), and platelet count(PLT). The reference intervals were determined by the nonparametric method. Cut-off values were determined by the receiver operating characteristics curve. A total of 135 healthy volunteers and 281 patients were enrolled. The 95% reference intervals were 96–195 s, 22–51 signal U/min, 〉1.6 for ACT, CR, PF respectively. In the 281 patients, the results of APTT, FIB, PLT, ACT, CR, and PF ranged from 20.5–300.0 s, 0.28–4.11 g/L,(19.0–387.3)×109/L, 80–514 s, 2.9–74 signal U/min, and 0.1–5.1 respectively. The cut-off values for transfusion were 〉208, ≤14, and ≤1.3 for ACT, CR, PF respectively. The cut-off values of Sonoclot analysis were within the manufacturer's reference intervals, while they were outside the reference intervals established in this study. The results suggested that the manufacturer's reference intervals were not suitable for Chinese. The reference intervals and cut-off values established in this study will be helpful to Chinese patients. Sonoclot analyzer has been widely used in many countries. But the reference intervals provided by the manufacturer were derived from only 45 participants, and there was no cut-off value for transfusion for Sonoclot analysis. This study aimed to establish reference intervals and transfusion criterion for Sonoclot analysis. Volunteers were recruited from healthy Chinese adults and patients undergoing cardiac surgery. Blood samples were withdrawn from forearm vein and measured for activated clotting time(ACT), clot rate(CR), platelet function(PF), activated partial thromboplastin time(APTT), fibrinogen concentration(FIB), and platelet count(PLT). The reference intervals were determined by the nonparametric method. Cut-off values were determined by the receiver operating characteristics curve. A total of 135 healthy volunteers and 281 patients were enrolled. The 95% reference intervals were 96–195 s, 22–51 signal U/min, 〉1.6 for ACT, CR, PF respectively. In the 281 patients, the results of APTT, FIB, PLT, ACT, CR, and PF ranged from 20.5–300.0 s, 0.28–4.11 g/L,(19.0–387.3)×109/L, 80–514 s, 2.9–74 signal U/min, and 0.1–5.1 respectively. The cut-off values for transfusion were 〉208, ≤14, and ≤1.3 for ACT, CR, PF respectively. The cut-off values of Sonoclot analysis were within the manufacturer's reference intervals, while they were outside the reference intervals established in this study. The results suggested that the manufacturer's reference intervals were not suitable for Chinese. The reference intervals and cut-off values established in this study will be helpful to Chinese patients.
出处 《Journal of Huazhong University of Science and Technology(Medical Sciences)》 SCIE CAS 2016年第4期614-617,共4页 华中科技大学学报(医学英德文版)
关键词 Sonoclot analyzer reference interval transfusion criterion Sonoclot analyzer reference interval transfusion criterion
  • 相关文献

参考文献11

  • 1Lee B, AI- Waili N, Butler G, et al. Assessment of heparin anticoagulation by Sonoclot Analyzer in arterial recon?struction surgery. Technol Health Care, 2011,19(19): 109- 114.
  • 2Schott U, Nilsson LG, Broman M, et al. Monitoring of low molecular weight heparin anticoagulation during haemodi?alysis with a Sonoclot Analyzer. Perfusion, 2010,25(4):191- 196.
  • 3Miyashita T, Kuro M. Evaluation of platelet function by sonoclot analysis compared with other hemostatic variables in cardiac surgery. Anesth Analg, 1998,87(6):1228-1233.
  • 4Chapin JW, Becker GL, Hulbert BJ, et al. Comparison of thromboelastograph and Sonoclot coagulation analyzer for assessing coagulation status during orthotopic liver trans?plantation. Transplant Proc, 1989,21(3):3539.
  • 5Bindi ML, Biancofiore GD, Consani G, et al. Blood coagu?lation monitoring during liver transplantation: Sonoclot analysis and laboratory tests. Minerva Anestesiol, 2001,67(5):359-369.
  • 6Sieders E, De Somer F, Bouchez S, et al. Haemostasis monitoring during sequential aortic valve replacement and liver transplantation. Acta Gastroenterol Belg, 2010,73(1): 65-68.
  • 7Nuttall GA, Oliver WC, Ereth MH, et al. Coagulation tests predict bleeding after cardiopulmonary bypass. J Cardio?thorac Vasc Anesth, 1997,11(7):815-823.
  • 8Yamada T, Katori N, Tanaka KA, et al. Impact of Sonoclot hemostasis analysis after cardiopulmonary bypass on post?operative hemorrhage in cardiac surgery. J Anesth, 2007,21(2): 148-152.
  • 9Sienco, Inc., Product Inserts. Available at: http://www.sie?nco.comlsonoclot-worldwide Iproduct-inserts.html. Acces?sed March 30, 2013.
  • 10Padhi S, Kemmis-Betty S, Rajesh S, et al. Blood transfu?sion: summary of NICE guidance. BMJ, 2015,351(2):395- 396.

同被引文献34

引证文献4

二级引证文献14

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部