Ocular biomechanical measurements on post- keratoplasty corneas using a Scheimpflug-based noncontact device 被引量:1

Ocular biomechanical measurements on post-keratoplasty corneas using a Scheimpflug-based noncontact device
原文传递
导出
摘要 AIM: To analyse ocular biomechanical properties, central corneal thickness (CCT) and intraocular pressure (lOP) in post-keratoplasty eyes, as compared to normal subjects, with a new Scheimpflug-based technology. Moreover, biomechanical data were correlated with the size and age of the donor and recipient corneas. METHODS: Measurements were conducted on 46 eyes of 46 healthy patients without any corneal pathology (age: 53.83 ±20.8y) and 30 eyes of 28 patients after penetrating keratoplasty (age: 49.43 ±21.34y). Ten biomechanical parameters, the CCT and lOP were recorded by corneal visualization Scheimpflug technology (CorVis ST) using high-speed Scheimpflug imaging. Keratometry values were also recorded using Pentacam HR system. Scheimpflug measurements were performed after 43.41± 40.17mo (range: 11-128mo) after the keratoplasty and after 7.64±2.34mo (range: 5-14mo) of suture removal. RESULTS: Regarding the device-specific biomechanical parameters, the highest concavity time and radius values showed a significant decrease between these two groups (P=0.01 and P 〈0.001). None of other biomechanical parameters disclosed a significant difference. The CCT showed a significant difference between post keratoplasty eyes as compared to normal subjects (P= 0.003) using the CorVis ST device. The lOP was within the normal range in both groups (P=0.84). There were no significant relationships between the keratometric data, the size of the donor and recipient, age of the donor and recipient and biomechanical properties obtained by CorVis ST. CONCLUSION: The ocular biomechanics remain stable after penetrating keratoplasty according to the CorVis ST measurements. Only two from the ten device-specific parameters have importance in the follow-up period after penetrating keratoplasty. AIM: To analyse ocular biomechanical properties, central corneal thickness (CCT) and intraocular pressure (lOP) in post-keratoplasty eyes, as compared to normal subjects, with a new Scheimpflug-based technology. Moreover, biomechanical data were correlated with the size and age of the donor and recipient corneas. METHODS: Measurements were conducted on 46 eyes of 46 healthy patients without any corneal pathology (age: 53.83 ±20.8y) and 30 eyes of 28 patients after penetrating keratoplasty (age: 49.43 ±21.34y). Ten biomechanical parameters, the CCT and lOP were recorded by corneal visualization Scheimpflug technology (CorVis ST) using high-speed Scheimpflug imaging. Keratometry values were also recorded using Pentacam HR system. Scheimpflug measurements were performed after 43.41± 40.17mo (range: 11-128mo) after the keratoplasty and after 7.64±2.34mo (range: 5-14mo) of suture removal. RESULTS: Regarding the device-specific biomechanical parameters, the highest concavity time and radius values showed a significant decrease between these two groups (P=0.01 and P 〈0.001). None of other biomechanical parameters disclosed a significant difference. The CCT showed a significant difference between post keratoplasty eyes as compared to normal subjects (P= 0.003) using the CorVis ST device. The lOP was within the normal range in both groups (P=0.84). There were no significant relationships between the keratometric data, the size of the donor and recipient, age of the donor and recipient and biomechanical properties obtained by CorVis ST. CONCLUSION: The ocular biomechanics remain stable after penetrating keratoplasty according to the CorVis ST measurements. Only two from the ten device-specific parameters have importance in the follow-up period after penetrating keratoplasty.
出处 《International Journal of Ophthalmology(English edition)》 SCIE CAS 2016年第2期235-238,共4页 国际眼科杂志(英文版)
关键词 corneal visualization Scheimpflugtechnology ocular biomechanics penetrating keratoplasty corneal visualization Scheimpflugtechnology ocular biomechanics penetrating keratoplasty
  • 相关文献

参考文献21

  • 1Soergel F, Jean B, Seiler T, Bende T, Mücke S, Pechhold W, Pels L. Dynamic mechanical spectroscopy of the cornea for measurement of its viscoelastic properties in vitro. Ger J Ophthalmol 1995;4(3):151-156.
  • 2Kynigopoulos M, Schlote T, Kotecha A, Tzamalis A, Pajic B, Haefliger I. Repeatability of intraocular pressure and corneal biomechanical properties measurements by the ocular response analyser. Klin Monbl Augenheilkd 2008;225(5):357-360.
  • 3Moreno-Montanes J, Maldonado MJ, Garcia N, Mendiluce L, Garcia-Gomez PJ, Segui-Gomez M. Reproducibility and clinical relevance of the ocular response analyzer in nonoperated eyes: corneal biomechanical and tonometric implications. Invest Ophthalmol Vis Sci 2008;49(3):968-974.
  • 4Chen S, Chen D, Wang J, Lu F, Wang Q, Qu J. Changes in ocular response analyzer parameters after LASIK. J Refract Surg 2010;26(4):279-288.
  • 5Ayala M. Corneal hysteresis in normal subjects and in patients with primary open-angle glaucoma and pseudoexfoliation Glaucoma. Ophthalmic Res 2011;46(4):187-191.
  • 6Greenstein SA, Fry KL, Hersh PS. In vivo biomechanical changes after corneal collagen cross-linking for keratoconus and corneal ectasia: 1-year analysis of a randomized, controlled, clinical trial. Cornea 2012;31(1):21-25.
  • 7Spoerl E, Terai N, Scholz F, Raiskup F, Pillunat LE. Detection of biomechanical changes after corneal cross-linking using Ocular Response Analyzer software. J Refract Surg 2011;27(6):452-457.
  • 8Shah S, Laiquzzaman M, Bhojwani R, Mantry S, Cunliffe I. Assessment of the biomechanical properties of the cornea with the ocular response analyzer in normal and keratoconic eyes. Invest Ophthalmol Vis Sci 2007;48(7):3026-3031.
  • 9Saad A, Lteif Y, Azan E, Gatinel D. Biomechanical properties of keratoconus suspect eyes. Invest Ophthalmol Vis Sci 2010;51(6):2912-2916.
  • 10Fontes BM, Ambrósio R Jr, Jardim D, Velarde GC, Nosé W. Corneal biomechanical metrics and anterior segment parameters in mild keratoconus. Ophthalmology 2010;117(4):673-679.

同被引文献1

引证文献1

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部