手脚污染测量仪手部探测器β表面活度响应的校准 被引量:5

Beta Surface Activity Response Calibration of Hand Detectors in a Hand-Foot Surface Contamination Monitor
在线阅读 下载PDF
导出
摘要 手脚表面污染测量仪用于测量辐射工作人员手部和脚部的放射性污染。为方便校准手部探头的表面活度响应,使校准结果更符合实际测量情况,制作了手部模型和90Sr+90Y薄膜放射源,分析研究了人体组织β散射对测量的影响,同时与使用铝衬底源的校准结果进行了比较分析。 Hand-foot surface contamination monitors are usually used in nuclear facilities for radiation protec- tion purpose, which measure surface contamination levels on workers' hands and feet. A hand model and two plastic film sources were made to facilitate the calibration of surface activity response of hand detectors in a monitor of this kind, and give rise to calibration results more in line with actual measurements. The effects of beta backscatter from hand tissue on measurement were investigated. The calibration results with aluminum substrate sources used were employed to perform a comparison in the meantime. Their difference was then ana- lyzed.
出处 《辐射防护通讯》 2015年第3期12-15,共4页 Radiation Protection Bulletin
关键词 手脚污染 表面污染 响应 校准 Hand-foot contamination Surface contamination measurement Response Calibration
  • 相关文献

参考文献8

  • 1中华人民共和国国家标准.电离辐射防护与辐射源安全基本标准[S].GB18871--2002.
  • 2IEC. Radiation Protection Instrumentation --Installed Per- sonnel Surface Contamination Monitoring Assemblies [ S ]. IEC 61098, 2003.
  • 3中华人民共和国国家标准.a、B表面污染测量仪与监测仪的校准[s].GB/T8997--2008.
  • 4陈子根 李星垣 帅小平.关于表面污染仪刻度的几个问题.核动力工程,1984,.
  • 5韩奎初,李景云.a表面污染监测仪的刻度[J].原子能科学技术,1979,(2):145.
  • 6张延生.大面积圆形探测器的刻度方法[J].核防护,1978,(4):40.
  • 7中华人民共和国国家标准.表面污染测定第1部分口发射体(EB,max>0.15MeV)和a发射体[s].GB/T14056.1~2008.
  • 8中华人民共和国国家计量检定规程.a、B和y表面污染仪[s].JJG478—1996.

共引文献1

同被引文献18

引证文献5

二级引证文献13

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部