螺旋CT扫描下新生儿性腺辐射防护干预的研究 被引量:4

Spiral CT Scanning Neonatal Gonadal Radiation Protection Intervention Under Study
原文传递
导出
摘要 目的用光致光测量技术检测头颅CT扫描下新生儿性腺防护衣内外辐射值以及防护器具对性腺的防护作用。方法在CT扫描时,将光致光测量芯片贴于防护器衣内外,分别测量其辐射剂量值,并进行内外对照。结果防护衣外剂量值为0.115 mGy,防护衣内剂量值为0.041 mGy。内外对照,新生儿性腺部位辐射剂量减低64%,经统计分析,辐射剂量值差异有统计学意义(P<0.05)。结论新生儿性腺部位不接受辅助防护的CT辐射剂量值有可能导致发生学的异常,辅助防护器具在相同条件下可以有效降低辐射剂量值。 Objective To measure, with light - induced opitieal measuring teehique, the interal and external radiation valcces to detect the protective effect of neonatal gonatal protection and the protective equipment under CT head scan. Methods With the photoluminescence measurements chip affixed to the inside and outside of neonatal gonadal protective equipment, radiation dose values were measured when CT scaing and the internal and external were compaired. Results The neonatal gonadal dose value of the unprotected parts is 0.115 mGy while the protective dose value of the protected pare is 0.041 mGy. Internal and external comparence shows 64% of the neonatal gonadal radiation dose was reduced when using the auxiliary protection, radiation dose was significant different (P 〈 0.05). Conclusion Neonatal gonadal site the values of the auxiliary CT radiation dose of the neonatal gonadal site may cause abnormal protection, The auxiliary protective device can effectively reduce the radiation dose values under the same condition.
出处 《中国辐射卫生》 2014年第4期332-334,共3页 Chinese Journal of Radiological Health
关键词 头颅CT扫描 新生儿性腺辐射剂量值 辐射防护干预 光致光测量技术 Neonatal Cranial CT Scan Radiation Dose Values Gonadal Radiation Protection Interventive Optically Stimulated Luminescence
  • 相关文献

参考文献13

  • 1United Nations Scientific Committee on the Effects of Atomic Radia- tion. Medical radiation exposures: UNSCEAR 2000 report to the General Assembly [ R ]. New York, NY : United Nations,2000 : 295 - 495.
  • 2Brenner D J, Hall EJ. Computed tomography an increasing source of radiation exposure [ J]. N Engl J Med ,2007,357:2277 -2284.
  • 3刘必跃.试述CT剂量的安全及管理[J].计量技术,2007(5):58-60. 被引量:4
  • 4Brenner DJ,Hall Elliston CD. Estimated radiation risks potentially associated with full -body CT screening [ J]. Radiology ,2004,232: 735 - 738.
  • 5Brenner DJ. Elliston CD,Hall E ,et al. Estimated risks of radiation induced fatal cancer from pediatric CT [ J ]. A JR,2001,176:289 - 296.
  • 6梁丽,赵蕊,李胜,岳强,徐健.X射线防护服所需铅当量的研究与应用[J].中国安全生产科学技术,2011,7(8):210-213. 被引量:7
  • 7朱大玲.螺旋CT低剂量扫描对早产儿颅脑病变的诊断与防护价值[J].中国妇幼保健,2011,26(34):5458-5459. 被引量:3
  • 8池田晟一.乳幼儿关节摄影には方'性腺防炉方案防炉具いつして[J].日本放射銭技木学会誌,1974:29(6):500.
  • 9Lawrence TD, Kevin AC,Yusuf EE, et al. Radiation dose reduction at a price:the effectiveness of a malegonadal shield during helical CT scans[ J]. BMC Medical . Imaging 2007,7:5.
  • 10UNSCEAR. Medical Radiation[ R]. Rxposure. 2008.

二级参考文献16

  • 1V.N.Filatov,许元巨.电离和电磁辐射防护材料的开发[J].产业用纺织品,1995,13(5):41-43. 被引量:4
  • 2张桂敏,郭建梅,周成飞,刘元理.医用射线防护屏蔽材料的研究及应用[J].中国辐射卫生,2005,14(3):184-185. 被引量:16
  • 3刘显坤,刘颖,唐杰,代君龙,胡春明.辐射屏蔽材料的研究进展[J].材料导报,2006,20(6):32-35. 被引量:30
  • 4郑均正,等.GBZl30-2002医用X射线诊断卫生防护标准[S].中华人民共和国卫生部,2002.
  • 5国际放射防护委员会第103号出版物[Z].北京:原子能出版社,2008.
  • 6苏森龄.X射线和射线防护手册[M].贵州:贵州人民出版社,1982.
  • 7国际电工委员会.医用诊断X射线辐射防护器具第3部分:防护服和性腺防护器具[S].国家药品监督管理局,2000,7.
  • 8Karmazyn B, Werner EA, Rejaie B, et al. Mesenteric lymph nodes in children: what is normal [J]? Pediatr Radiol, 2005, 35 (8): 774.
  • 9Naidich DP, Marshall CH, Gribbin C, et al. Low - dose CT of the lungs: preliminary observations [J] . Radiology, 1990, 175:729.
  • 10Brody AS, Frush DP, Huda W, et al. American academy of pediatrics section on radiology. Radiation risk to children from computed tomo- graphy [J] . Pediatrics, 2007, 120 (3): 677.

共引文献17

同被引文献30

引证文献4

二级引证文献17

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部