期刊文献+

对广西15家医院CR系统质量控制检测结果的分析 被引量:2

Research on the test results and analysis for the quality control of CR in Guangxi
暂未订购
导出
摘要 目的:通过了解、分析广西15家医院CR系统性能技术指标的合格情况,为采取相应的质量控制策略提供依据。方法:选取15台CR系统为检测和研究对象,根据2010年度卫生行业科研专项的实施方案确定性能指标,并依据国家标准对CR系统的性能进行检测和分析。结果:15台CR系统合格率仅为20%,CR专用检测项目主要不合格指标有:IP响应均匀性及一致性、IP响应线性、IP擦除完全性等。结论:加强CR系统的质量控制,定期对相关技术指标进行维护保养,从而提高医学影像诊断质量。 Objective: To understand the qualification of CR system in Guangxi, in order to take corresponding quality control strategy. Methods: Select 15 sets of CR system for detection and the research object, according to the implementation plan of 2010 annual health industry research to determine the performance index, and on the basis of national standards on the performance of CR system for testing and analysis. Results: The qualified rate of 15 sets CR system is only 20%, CR special detecting project main unqualified index have IP response uniformity and consistency, IP response linear, IP erasing completeness, etc. Conclusion: To strengthen the quality control of CR system, to maintain the relevant technical indexes on a regular basis, it is necessary to improve the quality of medical image diagnosis.
出处 《中国医学装备》 2015年第3期10-12,共3页 China Medical Equipment
基金 卫生行业科研专项(201002009)"辐射危害控制与核辐射卫生应急处置关键技术研究及其应用" 广西自然科学基金(2014GXNSFAA118223)"广西医用辐射危害评价与控制技术研究"
关键词 CR系统 质量控制 设备性能 维护保养 CR system Quality control Equipment performance Maintenance