不同级别迭代重建技术在颅脑CT扫描中的应用 被引量:7

Application of different level iterative reconstruction in CT examination of brain
暂未订购
导出
摘要 目的评价颅脑CT扫描中不同级别的迭代重建技术(SAFIRE)对图像质量和噪声的影响,探讨获得最佳图像质量的迭代重建级别。方法收集40例于本院行颅脑CT检查患者,分别采用滤波反投影法(FBP)及1-5级SAFIRE进行图像重建,比较不同组别图像基底节层面的组织CT值、噪声、信噪比、灰质可见度及图像对比度噪声比(CNR),并进行统计学分析。由两名影像学诊断医师采用双盲法以5分法对图像质量进行主观评价。结果 1SAFIRE 2级及3级图像评分值的差异无统计学意义,且二者图像评分最优,高于FBP图像。2SAFIRE 1-5级图像的噪声、信噪比及对比度噪声比均优于FBP图像,而灰质对比度差异均无统计学意义。结论在常规颅脑CT扫描中,与常规重建算法FBP比较,SAFIRE重建技术可提高图像质量,降低图像噪声。 Objective To evaluate the image quality and noise using different levels of iterative reconstruction technique (SAFIRE)on brain CT scan ,and to explore the best level iterative reconstruction which can provide the best image quality . Methods Siemens Somatom Definition Flash CT was used to perform brain CT examination on 40 patients .The original data of every case were reconstructed using traditional filtered back projection (FBP) and five different levels of iterative reconstruction techniques (SAFIRE1‐5)respectively .The image CT values ,noise ,GM visibility and contrast to noise ratio(CNR) of different levels of reconstruction images were analyzed statistically .The image quality was reviewed by two experienced doctors double-blindly using 5‐score system .Results ① There was no significant difference between SAFIRE2 and SAFIRE3 scores and the image quality were best .②The noise ,SNR and CNR .values of SAFIRE1-SAFIRE5 images were all better than those of FBP . While there was no significant difference in gray matter contrast .Conclusion On brain CT examination ,compared with FBP , the application of iterative reconstruction techniquecan significandy improve image quality and reduce image noise .
出处 《滨州医学院学报》 2015年第1期43-46,共4页 Journal of Binzhou Medical University
基金 滨州医学院科技计划(BY2011KJ055)
关键词 体层摄影术 X线计算机 辐射剂量 tomography X-ray computed radiation dose
  • 相关文献

参考文献10

  • 1Hara AK,Paden RG,Silva AC,et al.Iterative reconstruction technique for reducing body radiation dose at CT:feasibility study[J].AJR Am J Roentgenol,2009,193(3):764-771.
  • 2Mullins ME,Lev MH,Bove P,el al.Comparison of image quality between conventional and low-dose nonenhanced head CT[J].AJNR Am J Neuroradiol,2004,25(4):533-538.
  • 3Sagara Y,Hara AK,Pavlicek W,et al.Abdominal CT:comparision of low-dose CT with adaptive statistical iterative reconstruction and routine-dose CT with filtered back projection in 53patients[J].AJR Am J Roentgenol,2010,195(3):713-719.
  • 4Xu J,Mahesh M,Tsui BM.Is iterative reconstruction ready for MDCT[J].J Am Coll Radiol,2009,6(4):274-276.
  • 5陆秀良,曾蒙苏.迭代重建算法在CT中的应用[J].中国医疗设备,2012,27(4):128-131. 被引量:13
  • 6Schindera ST,Diedrichsen L,Müller HC,et al.Iterative reconstruction algorithm for abdominal multidetector CT at different tube voltages:assessment of diagnostic accuracy,image quality,and radiation dose in a phantom study[J].Radiology,2011,260(2):454-462.
  • 7贾楠,王新江,惠萍,李天文,孙红,李征芳.适应性统计迭代重建技术降低胸部CT扫描剂量的初步临床研究[J].中国医学影像学杂志,2010,18(6):551-553. 被引量:60
  • 8Moscariello A,Takx RA,Schoepf UJ,et al.Coronary CT angiography:image quality,diagnosetic accuracy,and potential for radiation dose reduction using a novel iterative image reconstruction technique-comparison with traditional filtered back projection[J].Eur Radiol,2011,21(10):2130-2138.
  • 9张宗锐,张永县,赵波,郭森林,刘云福,任悦,牛延涛,王振常.在颞骨MSCT检查中应用迭代算法:实验研究[J].中国医学影像技术,2012,28(5):1011-1014. 被引量:9
  • 10庄勋慧,冯卫华,徐文坚,曹会志.迭代重建技术在CT检查青年骶髂关节中的应用[J].中国医学影像技术,2012,28(4):796-799. 被引量:16

二级参考文献52

  • 1刘勇,张华.ALARA理论及CT低剂量应用研究进展[J].中国医学影像技术,2008,24(S1):17-20. 被引量:28
  • 2赵修义,李国英,陈静,肖红,刘汝勤.CT检查的防护探讨[J].中华放射医学与防护杂志,1995,15(1):57-57. 被引量:10
  • 3Marin D,Nelson RC,Schindera ST,et al.Low-tube-voltage,high-tube-current multidetector abdominal CT:improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm-initial clinical experience.Radiology,2010,254(1):145-153.
  • 4Hara AK,Paden RG,Silva AC,et al.Iterative reconstruction technique for reducing body radiation dose at CT:feasibility study.AJR Am J Roentgenol,2009,193(3):764-771.
  • 5Amis ES Jr,Butler PF Applegate KE,et al.American College of Radiology white paper on radiation dose in medicine.J Am Coll Radiol,2007,4(5):272-284.
  • 6Mulkens TH,Bellinck P,Baeyaert M,et al.Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations:clinical evaluation.Radiology,2005,237(1):213-223.
  • 7Lee CH,Coo JM,Ye H J,et al.Radiation dose modulation techniques in the multidetector CT era:from basics to practice.Radio Graphics,2008,28(5):1451-1459.
  • 8McCollough CH,Bruesewitz MR,Kofler JM Jr.CT dose reduction and dose management tools:overview of available potions.Radio Graphics,2006,26(2):503-512.
  • 9Kalra MK,Maher MM,Sahani DV,et al.Low dose CT of the abdomen:evaluation of image improvement with use of noise reduction filters pilot study.Radiology,2003,228(1):251-256.
  • 10Marin D,Nelson R C,Schindera S T,et al.Low-tube-voltage,high-tube-current multidetector abdominal CT:improved image quality and decreased radiation dose with adaptive statistical iterative reconstruction algorithm--initial clinical experience[J].Radiology,2010,254:145-153.

共引文献91

同被引文献34

引证文献7

二级引证文献30

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部