期刊文献+

UniCel DXI-800全自动微粒子化学发光分析仪测定血清黄体酮的性能评价 被引量:1

Performance evaluation of serum progesterone detecting by UniCel DXI-800 automated chemiluminescent analyzer
暂未订购
导出
摘要 目的:对UniCel DXI 800型全自动微粒子化学发光分析仪定量检测血清黄体酮(Prog)进行性能评价。方法运用该系统检测血清Prog,并对其灵敏度、精密度、准确度及特异性等进行方法学评价。同时与Access 2免疫分析系统进行相关性比较。结果该系统检测血清 Prog 的精密度较高,变异系数小于10%,最低检测限为0.1 ng/mL,回收率范围为95.3%∽104.7%,抗干扰能力各项指标均达到较高水准。与Access 2化学发光免疫系统结果比较具有很好的相关性,得到回归方程:Y=0.84X-0.63,r=0.978。结论该系统对定量检测血清Prog 有较高的灵敏度和高特异性,且精密度符合临床要求,与Access 2检测系统有很好的相关性,值得推广。 Objective To evaluate the quantitative detection of serum progesterone by UnicellDXI-800 automated chemilumines-cent analyzer.Methods To detect the serum progesterone,and evaluate the sensitivity,precision,accuracy and specificity.At the same time,the correlation with Access 2 immune system was analyzed as well.Results The precision of this system of serum pro-gesterone detection was high,with the variation coefficient less than 10%,the sensitivity of 0.1 ng/mL,and the recovery range of 95.3% to 104.7%.And the anti-jamming capability indexes reached higher level.It had good correlation compareing with the pre-vious Access 2 chemiluminescence immune system,and the regression equation obtained wasY= 0.84X-0.63,r= 0.978.Conclu-sion The system for quantitative detection of serum progesterone had high sensitivity and specificity,and its precision could consistent with clinical requirements.It has good correlation with Access 2 chemiluminescence immune system and worthy of promotion.
出处 《国际检验医学杂志》 CAS 2014年第12期1628-1628,I0001,共2页 International Journal of Laboratory Medicine
关键词 黄体酮 准确度 精密度 变异系数 全自动微粒子化学发光分析仪 progesterone accuracy precision coefficient of variation automated chemiluminescent analyzer
  • 相关文献

参考文献12

二级参考文献24

  • 1陈悦洲,李建华,郑行.激素与早期胚胎发育[J].生殖与避孕,2007,27(3):209-212. 被引量:4
  • 2王淑贞 严仁英.实用妇产科学.胎儿附属物的形成及其功能(第2版)[M].人民卫生出版社,1964,10.43-51.
  • 3张以文.女性生殖生理[A].见:曹泽毅主编.中华妇产科学:第1版[C].北京:人民卫生出版社,1999.36-44.
  • 4U. S. Department of Health and Human Services.Medicare, Medicaid and CLIA program: Regulations implementing the Clinical Laboratory Improvement Amendment of 1988(CLIA)[S]. Final rule. Fed Regist, 1992,57(40): 7002-7186.
  • 5Kolibianakis EM, Zikopoulos K, Smitz J, et al. Elevated progesterone at initiation of stimulation is associated with a lower ongoing pregnancy rate after IVF using GnRH antagonists [J]. Hum Reprod,2004,19(7): 1525-1529.
  • 6Segal S, Glatsein I, Mcshane P, et al. Premature luteinization and in vitro fertilization outcome in gonadotropin/gonadotropin-releasing hormone antagonist cycles in women with polycystic ovary syndrome [J]. Fertil Steril,2008,4(2): 1-5.
  • 7Venetis CA, Kolibianakis EM, Papanikolaou E, et al. Is progesterone elevation on the day of human chorionic gonadotrophin administration associated with the probabilty of pregnancy in vitro fetriliza-tion: a systematic review and meta-analysis [J]. Hum Reprod Update,2007,13(4): 343-355.
  • 8Amui J, Check JH, Choe JK, et al. Relationship of the serum progesterone (p) level the day after human chorioinc gonadotropin (hCG) injection on outcome following in vitro fertilization-embryo transfer (IVF-ET) [J]. Fertil Steril,2006,85(Suppl 2):S15.
  • 9[1]俞霭峰.妇产科内分泌学.上海科学技术出版社,1982:44.
  • 10Gardner DK, Weissman A, Howles CM, et al. Textbook of assisted reproductive techniques laboratory and clinical perspectives. United Kingdom: Martin Dunitz, 2001,468-469.