摘要
目的 选择适用于医用乳胶手套辐射灭菌验证剂量的最隹的有效的运行模式。方法 采用不同的运行模式对医用乳胶手套进行辐射灭菌 ,测量在辐照吊箱中该产品的剂量分布。结果 获得不同运行模式下辐照吊箱中医用乳胶手套的剂量分布。结论 采用源位标准高度 ,在辐射灭菌过程中对辐照吊箱中的产品进行翻箱 ,达到规定的停顿时间 ,即可使实施的剂量符合该产品辐射灭菌验证剂量要求 ,且此运行模式为最隹。
Objective To determine the optimal radiation mode for sterilization of surgical latex gloves. Methods The dose distribution in the product loads inside radiation container was measured in different radiation mode.Results Data of dose distribution in the product load in different radiation mode were obtained. Conclusion At pre-set height of source working position and within pre-determined dwelling time of each irradiation container staying at the irradiation position, the delivered dose can meet the customers requirement by the radiation mode that, in the half cycle of radiation process, turns the horizontal middle layers of the product to the upper or lower layers, and the upper and lower layers to the middle layers.
出处
《苏州大学学报(医学版)》
CAS
2002年第3期244-246,共3页
Suzhou University Journal of Medical Science
关键词
医用乳胶手套
辐射灭菌
验证剂量
surgical latex glove
radiation sterilization,
verification dose