摘要 根据反馈原理,采用乘法器设计和实现了一个用于电磁感应阻抗测量的高精度鉴相器。当激励频率为 50KHz时,鉴相器的相位灵敏度△Vq/△φ=11.46V/rad;输出端噪声的标准差 STD(△φ[rad])=1.42×10-4rad;输出端的信噪比 SNR=41.842dB;系统响应时间为 3s。 The high-resolution phase detector for impedance measurement by electromagnetic induction, based on the principle of feedback, is electrically realized with multipliers. When the excitation frequency is 50KHz, the sensitivity, the STD (standard deviation) of the measured phase, the SNR (signal noise ratio) and the response time of the system are 11.46V/rad, 1.42*10-4rad, 41.842dB and 3s respectively.
出处 《医疗卫生装备》 CAS 2002年第3期18-19,22,共3页 Chinese Medical Equipment Journal
关键词 电阻抗 电磁感应 鉴相器 impedance electromagnetic induction phase detector
  • 相关文献

参考文献5

  • 1[1]Johannes Netz,Ewald Forner,sabine Haagemann. Contactless impedance measurement by magnetic induction. Physiol.Meas.,
  • 2[2]N.G.Gencer and M.N.Tek.Electrical conductivity imaging via con tactless measurements.IEEE Trans.Med.Imag.,1999,18(7):617~627
  • 3[3]H. Griffiths,W.r.Stewart,W.Gough.Magnetic Induction Tomography.Ann N Y Acad.Sci,1999,873:335~345
  • 4[4]A.V.Korjenevsky and A.V.Cherepenin.Progress in realization of magnetic induction tomography.Ann N Y Acad Sci,1999,873:346~352
  • 5[5]H.Scharfette,,H.K.Lackner,J.Rosell,High resolution hardware for magnetic induction tomography (MIT)at low frequencies.Biomedical application of EIT,the 2nd EPSRC Engineering Network meeting, UCL;71~75