期刊文献+

火焰原子吸收法测定和计算误差的减小 被引量:1

Minimizing the Detection and Calculation Error in FAAS Analysis
在线阅读 下载PDF
导出
摘要 通过对火焰原子吸收法样品分析误差产生的各主要环节的分析和实验的验证 ,指出在盐湖资源开发过程产生的混合样品分析中 ,样品稀释倍数、光谱干扰、电离干扰和计算方法是影响结果准确性的主要因素 ,进一步提出了减小误差的办法 ,并给出了用Excel处理数据的示例表格和相应的VBA程序代码。 s:By theoretical analysis and experimental data comparison, this paper notifies that the error in FAAS analysis for the exploitation of salt-lake resources is formed in dilution stage, detection stage and calculation stage, and that it is mainly determined by the combination of the dilution factor, spectral interference, ionization interference, and the calculation method. In order to reduce the error to a bearable degree for the development of chemical processes, the author suggests a best combination of the three stages and offers a solution in the calculation process in the form of excel spreadsheet, with the supporting VBA code given.
出处 《盐湖研究》 CSCD 2002年第2期40-46,共7页 Journal of Salt Lake Research
关键词 火焰原子吸收法 干扰 线性最小二乘法 拟合 VBA EXCEL FAAS FAAS Interference Linear least squares method Quadratic fitting method VBA
  • 相关文献

参考文献2

  • 1北京第二光学仪器厂.原子吸收分光光度计或研发分析手册[M].,1990..
  • 2程光钺 邝平先.化学中的数学模型方法[M].北京:科学出版社,2000..

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部