期刊文献+

RIA与GICA测定尿HCG的结果比较

Comparision of Results of GICA and RIA in Determination of HCG in Urine
暂未订购
导出
摘要 目的 :对胶体金免疫层析试验与放射免疫分析随机尿HCG的结果进行比较分析。方法 :采用胶体金免疫层析测定试条和放射免疫分析法对 1 4 8例育龄妇女随机尿HCG进行测定。结果 :胶体金免疫层析试条对随机尿HCG的灵敏度为 1 2 0 0mIU/ml,检测范围在 1 2 0~ 42 7570mIU/ml,当尿HCG浓度大于 42 7570mIU/ml时 ,该试条出现带现象 ,放射免疫分析法的灵敏度为 2 5mIU/ml,没有带现象。结论 :胶体金免疫层析试条具有操作简单、快速等优点 ,与放射免疫分析法相比 ,存在灵敏度低、带现象和不能定量等问题 ,在日常应用中应加以注意。 Objective To compare the results of golden immunochromastographic assay (GICA) and radioimmunoassay (RIA) in determination of human chorionic gonadotropin (HCG) in urine. Methods 148 random urine specimens from child-bearing age women were determined for HCG with both GICA and RIA. Results The lowest detactable content of HCG in urine was 120.0mIU/ml with GICA and the detectable range of concantranion was 120-427570mIU/ml, over which the zone phenomena (antigen excess) occurred with false negative rasult. The sensitivity with RIA was 25mIU/ml and there was no zone phenomena. Conclusion The GICA wasa simple and rapid method in determination of HCG in urine, but there was the disadvantoge of lower sensitivity and presence of zone phenomena.
出处 《放射免疫学杂志》 CAS 2002年第1期8-9,共2页 Journal of Radioimmanology
关键词 胶体金免疫层析 放射免疫分析 绒毛膜促性腺激素 尿液 Urine, human chorionic gonadotropin (HCG), radioimmunoassay (RIA), golden immunochromatographic assay (GICA)
  • 相关文献

参考文献8

二级参考文献6

  • 1杨振修,陈宪莲,陶义训.银强化金标记免疫吸附剂试验的方法学研究[J].上海免疫学杂志,1996,16(1):42-43. 被引量:40
  • 2王张琼,临床医学,1986年,6卷,4期,216页
  • 3Fan D F,Clin Chem,1985年,31卷,494页
  • 4余秋生,中日友好医院学报,1992年,6卷,3期,145页
  • 5周仪如,蚌埠医学院学报,1987年,12卷,1期,29页
  • 6武建国,实用临床免疫学检验,1989年,13页

共引文献91