期刊文献+

SLICPS2000型裂隙灯图像分析系统可重复性的研究 被引量:1

A study on SLICPS 2000 slit lamp image analysis system
暂未订购
导出
摘要 目的 检测SLICPS 2 0 0 0型裂隙灯图像分析系统在白内障定量分析方面的可重复性。方法  3名检查者分别对同一组大鼠半乳糖性白内障晶状体图像进行定量分析 ;对同一只先天性白内障眼进行 1 0次分析 ;对同一幅白内障图像分析 1 0次。结果 不同检查者对同一组病例的检测 ,对同一病例进行多次检测 ,对同一图像多次分析 ,其检测结果比较差异均无显著性。结论 SLICPS 2 0 0 0型裂隙灯图像分析系统的白内障定量分析有很好的可重复性。 OBJECTIVE To test repeat performance of SLICPS 2000 slit lamp image in quantitative analysis of cataract.METHODS Three operators performed quantitative determinations of lens images separately in same one group of rat with galactose-cataract. Lens was determined at 10 times in one eye with congenital cataract. Each frame of the images was analyzed at 10 times.RESULTS There were no statistically significant differences of values among three groups.CONCLUSION SLICPS 2000 slit lamp image collect and process system has favorable repeat performance, can be used for quantitative diagnose of cataract.
出处 《中国中医眼科杂志》 2001年第4期188-190,共3页 China Journal of Chinese Ophthalmology
基金 国家自然科学基金 (3970 70 8)
关键词 裂隙灯 图像分析 晶状体 可重复性 slit lamp image analysis lens
  • 相关文献

参考文献4

  • 1Sasaki K, Sakamoto Y, Shibata T, et al. The multi - purpose camera: A new anterior eye segment analysis system [J] . Ophthalmic Res, 1990, 22 (Supp1 1): 3-8.
  • 2Sparrow J M, Brown N A P, Shun - shin G A, et al. The Oxford modular cataract image analysis system [ J ] . Eye, 1990, 4:638 - 648.
  • 3Vivino M A, Chintalagiri S, Trus B, et al. Development of a scheimpflug slit lamp caraera system for quantitative densitometric analysis [J] .Eye, 1993, 7:791-798.
  • 4马文新,唐由之.晶体混浊分期分级的研究现状[J].中国中医眼科杂志,1997,7(2):120-122. 被引量:6

二级参考文献3

  • 1John M. Sparrow FRCS,William Ayliffe,Anthony J. Bron,Nicholas P. Brown,Adrian R. Hill. Inter-observer and intra-observer variability of the Oxford clinical cataract classification and grading system[J] 1988,International Ophthalmology(3):151~157
  • 2J. M. Sparrow F.R.C.S.,A. J. Bron,N. A. P. Brown,W. Ayliffe,A. R. Hill. The Oxford Clinical Cataract Classification and Grading System[J] 1986,International Ophthalmology(4):207~225
  • 3高文勇,商福.无损伤的白内障定量检查[J].国外医学(眼科学分册),1991,15(3):160-163. 被引量:2

共引文献5

同被引文献4

  • 1黄庆玲 胡天圣 毛文书.人晶体可溶性高分子量蛋白的分离和测定[J].中华眼科杂志,1984,20(5):257-261.
  • 2Delaye M,Tardieu A.Short-range order of crystallin proteins accounts for eye lens transparency[J].Nature 1983;302:415-417.
  • 3Sun TX,Das BK,Liang JJ.Conformational and functional differences between recombinant human lens alpha A- and alpha B-crystallin[J].J Biol Chem 1997;272(10):6220-6225.
  • 4Kim YH,Kapfer DM,Boekhorst J,Lubsen NH.Deamidation,but not truncation,decreases the urea stability of a lens structural protein,betaB1-crystallin[J].Biochemistry 2002;41(47):14076-14084.

引证文献1

二级引证文献12