摘要 研究了 ICP光源中试液进样量对稀土元素谱线强度、背景强度 ,谱线强度测量的相对标准偏差 ,元素检出限 ,发光效率及试样相对利用率的影响。结果表明 ,降低试液进样量对稀土元素谱线强度有轻度影响 ,并不影响测量的相对标准偏差和元素的检出限。降低试液进样量将提高发光效率及试样利用率 。 Effects of sample uptake rate on spectral line intensities, background intensities, relative standard deviations of line to background ratio, detection limits, signal emitted rate and sample relative utilization rate were studied. The results showed that the effect of sample uptake rate on spectral line intensities was slight. Sample uptake rate did not affect RSD and detection limits. Lower sample uptake rate could increase signal emitted rate and sample realtive utilization rate. This method can be applied to the rare metals, precious materials and radioactive elements.
出处 《分析科学学报》 CAS CSCD 北大核心 2001年第5期391-394,共4页 Journal of Analytical Science
关键词 ICP-AES 试液进样量 稀土元素 电感耦合 等离子体原子发射光谱 分析 ICP AES Sample uptake rate Analytical performance Rare earth elements
  • 相关文献

参考文献1

二级参考文献3

  • 1Wing R K,,Pererson V J,Fassel V A. Applied Spectroscopy . 1979
  • 2l Sweedler T v,Bihon A B,Epperson P M,Sims G R,Deton M B. Analytical Biochemistry . 1988
  • 3Sweedler J W,Deton M B. Applied Spectroscopy . 1987