摘要 目的对JZ-1C型二氧化碳(CO2)激光治疗仪指示光功率、时间稳定度、复现性及示值相对偏差等指标进行检测,以保证CO2激光治疗仪的临床安全,提高其临床治疗效果。方法采用AD7888模数转换器对上述各项参数进行检测。结果该治疗仪各项参数值均符合国家强制标准要求,可以继续应用于临床。结论对CO2激光治疗仪进行有效地质量控制,可以有效规避临床风险,提高疗效。 Objective To improve the clinical safety and clinical effectiveness of JZ-1C CO2 laser treatment instrument by measuring its parameters including the power of indicator light, time stability, reproducibility and relative deviation of indicating value ((3p). Methods Measured the above parameters by using AD7888 analog-digital converter. Results The measured values of the above parameters can meet the requirements of national mandatory standards in China so that the JZ-1C CO2 laser treatment instrument can be used in clinical sequentially. Conclusion The effective quality control of CO2 laser treatment instrument should be executed to avoid the clinical risks and improve the clinical effectiveness.
作者 陈文辉 高原
出处 《中国医疗设备》 2014年第3期63-65,150,共4页 China Medical Devices
关键词 激光治疗仪 质量控制 光功率 时间稳定度 复现性 示值相对偏差 laser treatment instrument quality control the power of indicator light time stability reproducibility relative deviation of indicating value
  • 相关文献

参考文献12

二级参考文献79

共引文献35

同被引文献35

  • 1刘华绪,任秋实.长脉宽1064nm Nd:YAG激光治疗血管性皮肤病的原理及其应用[J].中国麻风皮肤病杂志,2006,22(7):588-591. 被引量:15
  • 2苏志深.901-C型二氧化碳激光治疗机故障检修4例[J].医疗设备信息,2006,21(8):111-111. 被引量:1
  • 3朱平,吴小光.激光与激光医学[M].北京:人民军医出版社,2011:29-56.
  • 4陈家璧,彭润玲.激光原理及应用[M].北京:电子工业出版社,2013.
  • 5YY0789-2010,Q开关Nd:YAG激光眼科治疗机[S].
  • 6YY0307-2011,连续波掺钕钇铝石榴石激光治疗机[S].
  • 7YY0846-2011,激光治疗设备掺钬钇铝石榴石激光治疗机[S].
  • 8YY0845-2011,激光治疗设备半导体激光光动力治疗机[S].
  • 9GB12257-2000,氦氖激光治疗机通用技术条件[S].
  • 10GB11748-2005,二氧化碳激光治疗机[S].

引证文献5

二级引证文献15

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部