XT-4000i多功能全自动血细胞分析仪体液模式性能验证 被引量:5

Evaluation the performance of the body fluid mode on the automatic analysis system-XT-4000i
暂未订购
导出
摘要 目的:评价日本SYSMEX生产的XT-4000i多功能全自动血细胞分析仪(以下简称XT-4000i分析仪)体液模式的性能。方法采用XT-4000i分析仪分别检测体液模式下各项参数(WBC和RBC )的精密度、准确度、携带污染、仪器测定线性验证。结果 WBC-BF精密度为0.90%~5.54%,RBC-BF精密度为1.13%~5.13%。携带污染率WBC、RBC分别为0.09%、0.14%;仪器有较宽的WBC和RBC的检测线性;仪器测定与手工镜检的比较结果均满足要求。结论 XT-4000i分析仪体液模式各性能均符合仪器的要求范围,可用于临床体液分析检测。 Objective To evaluate the performance of the body fluid mode on the automatic analysis system-XT-4000i .Methods XT-4000i analyzer was used to detect the precision ,accuracy ,carryover rate ,meanwhile to validate the detection limit of the pa-rameters(WBC and RBC) .Results The range of the precision for WBC-BF was 0 .90% -5 .54% ,for RBC-BF was 1 .13% -5 .13% .The carry over rates were 0 .09% for WBC and 0 .14% for RBC ,respectively .As for as the analyzer range ,the results of the linearity and the detection limit verification were in line with the requirements of the analyzer ;the comparison between the man-ual microscopy results and the analyzer results meet the requirements of XT-4000i analyzer .Conclusion The performance of XT-4000i analyzer is in line with the requirements of the instrument range ,which can be used for clinical body fluid analysis and detection .
作者 王梅 吴燕
出处 《重庆医学》 CAS CSCD 北大核心 2013年第33期4002-4003,4006,共3页 Chongqing medicine
基金 国家临床重点专科建设项目经费资助[财社(2010)305号]
关键词 分析仪 性能评价 精密度 准确度 携带物染率 automatic analyzer precision accuracy carryover clinical application
  • 相关文献

参考文献11

二级参考文献24

共引文献71

同被引文献41

  • 1凌晓午,刘莲,安邦权,周湘红,夏世勤,叶艾竹,蒋亚风.Bayer ADVIA120全自动血细胞分析仪性能评价[J].检验医学与临床,2005,2(4):184-186. 被引量:8
  • 2张莉,吴炯,郭玮,潘柏申.医学检验检测系统应用前的性能评价[J].检验医学,2006,21(6):560-563. 被引量:71
  • 3毕波,吕元.定量检测方法学性能验证的系统设计[J].中华检验医学杂志,2007,30(2):143-145. 被引量:154
  • 4Elkin Simson,崔巍.全自动血液分析仪的复检标准[J].中华检验医学杂志,2007,30(4):371-373. 被引量:21
  • 5杨有业,张秀明.临床检验方法学评价[M].北京:人民卫生出版社,2008:118-127.
  • 6叶应抚,王毓三,申子瑜.全国临床检验操作规程[M].3版.南京:东南大学出版社,2006:180-182.
  • 7CLSI NCCLS EP5-A2.Evaluation of precision performance of quantitative measurement methods[S].CLSI Approved guideline,2004.
  • 8CLSI NCCLS H26-A2.Validation,verification and quality assurance of automated hematology analyzers[S].CLSI Approved standard,2010.
  • 9CLSI NCCLS EP6-A.Evaluation of the linearity of quantitative measurement procedures:astatistical approach[S].CLSI Approved guideline,2003.
  • 10Takemura H,Tabe Y,Ishii K,et al.Evaluation of capability of cell count and detection of tumor cells in cerebrospinal and boby fluids by automated hematology analyzer[J].Rinsho Byori,2010,58(6):559-564.

引证文献5

二级引证文献19

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部