期刊文献+

一种高性能流动体系电流检测器 被引量:1

A High-Performance Amperometric Detector for Flow-through Systems
在线阅读 下载PDF
导出
摘要 本文提出了一种高性能流动体系电流检测器,它采用薄层型结构,工作电极与辅助电极面对面放置,参比电极放在工作电极的下游,用一片离子选择膜将工作电极与辅助电极隔开。这样,不仅克服了目前多数电流检测器所存在的工作电极表面各点电位不一致。电流密度分布不均匀,溶液欧姆电位降大,工作电极面积利用率低等缺点,而且保持了原有检测器中工作电极与辅助电极处于上、下游,避免了辅助电极反应产物的干扰的特点,提高了检测器的灵敏度,扩大了浓度线性范围。 A high performance amperometric detector for flow-through systems is presented. The detector takes the structure of thin layer cell with working electrode(WE), opposite auxiliary electrode(AE) and the reference electrode(RE) at the downstream of WE. A piece of ion-exchange membrane is used to keep WE from the effect of AE products. This amperometric detector not only maintains the advantages of conventional detectors, but also overcomes their drawbacks. The main characteristics are as follows: iden ical potential at different position of WE; uniform current distribution of WE surface; low ohmic potential drop; high utilization of WE area; no influence from AE; high sensitivity; good current linearity for broad range of concentration.
机构地区 厦门大学化学系
出处 《分析化学》 SCIE EI CAS CSCD 北大核心 1991年第2期253-256,共4页 Chinese Journal of Analytical Chemistry
基金 国家自然科学基金
关键词 电解池 薄层 流动体系 电流检测器 Flow-through electrochemical detector, Amperometric detector, Thin layer cell
  • 相关文献

参考文献3

  • 1包立源,厦门大学学报,1989年,28卷,4期,381页
  • 2包立源,厦门大学学报,1988年,27卷,2期,191页
  • 3Kisshger P T,分析化学,1986年,14卷,1期,74页

同被引文献17

  • 1Lin TsuKeng, Chen YenChen, Tien ChaoWu. Autocatalyric Electroreduction of Iodate in Acidic Solution. Scientia Sirtica, 1966,15:792.
  • 2田昭武 陈体衔 林仲华等.DHZ-1型电化学综合测试仪[J].厦门大学学报,1978,17:150-150.
  • 3田昭武 蔡加勒 邱贞花等.DD-1型电镀参数测试仪[J].厦门大学学报,1979,18:68-68.
  • 4Tian Z W. Polarization Theory of Porous Electrode-Characteristic Currents for Transport inside Porous E lectrode. Scientia Sinica, 1981,24 : 1103.
  • 5Tian Z W, Lin Zugeng, You Jinkua. Polarization Theory of Porous Electrode Uneven Liquid Film Model for Gas Diffusdion Porous Electrode. Scientia Sinica, 1981,24: 1391.
  • 6Tian Z W, R. Z. Hu, H. S. Lin, et al. Effect of Suppression Efficiency on Sensitivity in Ion Chromatography. Journal of Chromatography, 1988,439 : 151157.
  • 7Tian Z W, R. Z. Hu, H. S. Lin, et al. HighPerformanee Electrochemical Suppressor for Ion Chromatography. Journal of Chromatography, 1988,439 : 159-163.
  • 8You J K, Lin Z G,Tian Z W. High-Performance Gas Diffusion Porous Electrode Starved of Electrolyte Solution. Chinese Chemical Letters, 1990,1 : 151-160.
  • 9Bao L Y, Tian Z. W. A New Approach To Coulometry for Flow Through Systems. J. Electroanalysis, 1991,3 : 49-51.
  • 10Tian Z W, X. D. Zhuo, J. Q. Mu, et al. A New Electrochemical Scanning Tunneling Microscope. Ultramicros- copy, 1992,42-44, :460-463.