Osstell ISQ^(TM)监测种植体稳定性的临床研究 被引量:22

The predictive value of RFA measurements using Osstell ISQ^(TM) in implant stability determination: a clinical study
暂未订购
导出
摘要 目的:分析种植体初期稳定系数(ISQi)与植入扭矩值(ITv)在评估初期稳定性中的关系;分析不同初期稳定性种植体在骨愈合期种植体稳定系数(ISQ)的变化趋势。方法:选择牙缺失患者34例,共55枚种植体。应用Osstell ISQTM共振频率分析仪于种植体植入时测量ISQ和记录ITv,并于植入后4周、8周、12周监测ISQ变化。结果:ISQi与ITv存在强正相关性。低初期稳定性组种植体植入后ISQ呈持续上升过程,高初期稳定性组ISQ则呈先降后升趋势,4周降至最低点。结论:Osstell ISQTM能良好评价和监测种植体稳定性,ISQi与ITv均为评价种植体初期稳定性的客观指标,种植体稳定性的变化趋势与初期稳定性范围相关。 Objective: The purpose of the present study was to analyze the relationship between initial im-plant stability quotients (ISQi) and insertion torque value (ITv) on implant stability evaluatation, and discuss the change of implant stability quotients (ISQ) of different initial stability over the healing time. Methods: Thirty-four patients who needed one or more dental implants were collected in the research. A total of 55 implants were inserted. At the time of implant placement, each implant was tested for pri-mary stability by resonance frequency analysis (RFA) and insertion torque. RFA was performed again during examinations at 4, 8 and 12 weeks following surgery. Results: A strong, positive correlation was found between ISQi and ITv. In the lower initial stability group, the ISQ steadily increased throughout the healing period, whereas in the higher groups decreased in the first four weeks and then increased gently. Conclusions: Osstell ISQTM is a quite sensitive simple-used instrument to measure and evaluate the implant stability. ISQi and ITv are the favorable indices to evaluate the initial implant stability. ISQ change pattern is closely related to the primary stability.
出处 《中国口腔种植学杂志》 2013年第3期133-137,共5页 Chinese Journal of Oral Implantology
基金 2010年广东省科技计划项目(2010B031100007)
关键词 共振频率分析 种植体稳定性 种植体稳定性系数 植入扭矩 resonance frequency analysis implant stability implant stability quotients insertion torque value
  • 相关文献

参考文献16

  • 1Balshi S F, Allen F D, Wolfinger G J, et al. A reso- nance frequency analysis assessment of maxillary and mandibular immediately loaded implants. Int J Oral Maxillofac Implants,2005,20 (4) :584-594.
  • 2A1-Nawas B, Hangen U, Duschner H, et al. Turned, machined versus double-etched dental implants in vivo. Clin Implant Dent Relat Res,2007,9 (2) :71- 78.
  • 3周炜,宋应亮.共振频率分析在口腔种植稳定性测量中的应用[J].国际生物医学工程杂志,2006,29(2):120-123. 被引量:7
  • 4Turkyilmaz I, Aksoy U, Mcglumphy E A. Two alter- native surgical techniques for enhancing primary implant stability in the posterior maxilla: a clinical study including bone density, insertion torque, and resonance frequency analysis data. Clin Implant Dent Relat Res,2008,10 (4) :231-237.
  • 5Rabel A, Kohler S G, Schmidt-Westhausen A M. Clinical study on the primary stability of two dental implant systems with resonance frequency analysis. Clin Oral Investig,2007,11 (3) :257-265.
  • 6Da C H, Francischone C E, Filbo H N, et al. A comparison between cutting torque and resonance frequency in the assessment of primary stability and final torque capacity of standard and TiUnite single-tooth implants under immediate loading. Int J Oral Maxillofac Implants,2004,19 (4) :578-585.
  • 7Akca K, Chang T L, Tekdemir I, et al. Biomechani-cal aspects of initial intraosseous stability and im- plant design: a quantitative micro-morpbometric analysis. Clin Oral Implants Res,2006,17 (4) :465- 472.
  • 8Mesa F, Munoz R, Noguerol B, et al. Multivariate study of factors influencing primary dental implant stability. Clin Oral Implants Res,2008,19 (2) :196- 200.
  • 9Tozum T F, Turkyilmaz I, Mcglumphy E A. Rela- tionship between dental implant stability determined by resonance frequency analysis measurements and peri-implant vertical defects: an in vitro study. J O- ral Rehabil,2008,35 (10) :739-744.
  • 10Raghavendra S, Wood M C, Taylor T D. Early wound healing around endosseous implants: a review of the literature. Int J Oral Maxillofac Implants, 2005,20 (3) :425-431.

二级参考文献21

  • 1Meredith N,Alleyne D,Cawley P.Quantitative measurement of the stability of the implant-tissue interface using resonance frequency analysis[J].Clin Oral Impl Res,1996,7:261-267.
  • 2Nedir R,Bischof M,Szmukler-Moncler S,et al.Predicting ossointegration by means of implant primary stability:A resonance frequence nalysis study with delayed and immediately loaded ITI SLA implants[J].Clin Oral Impl Res,2004,15:520-528.
  • 3Huang HM,Chiu CL,Yeh CY,et al.Factors influencing the resonance frequency of dental implants[J].J Oral Maxillofac Surg,2003,61(10):1184-1188.
  • 4Sennerby L,Meredith N.Resonance frequency analysis:Measuring implant stability and osseointegration[J].Compend Contin Educ Dent,1998,19(5):493-502.
  • 5Roland G,Lars S,Neil M.Resonance frequency analysis of implants subjected to immediate or early functional occlusal loading[J].Clin Oral Impl Res,2004,15(4):428-434.
  • 6Horwitz J,Zuabi O,Peled M.Resonance frequency analysis in immediate loading of dental implants[J].Refuat Hapeh Vehashinayim,2003,20(3):80-88.
  • 7Meredith N,Rasmusson L,Sennerby L.Mapping implant stability by resonance frequency analysis[J].Med Sci Res,1996,24:191-193.
  • 8Meredith N.Assessment of implant stability as a prognostic determinant[J].Clin Oral Impl Res,1997,8:234-243.
  • 9Huang HM,Chiu CL,Yeh CY,et al.Early detection of implant healing process using resonance frequency analysis[J].Clin Oral Impl Res,2003,14(4):437-443.
  • 10Balleri P,Cozzolino A,Ghelli L,et al.Stability measurements of osseo-integrated implants using Osstell in partially edentulous jaws after 1 year of loading:A pilot study[J].Clin Impl Dent Relat Res,2002,4(3):128-132.

共引文献6

同被引文献122

引证文献22

二级引证文献111

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部