期刊文献+

氢化物发生-四通道原子荧光光谱法同时测定高纯金中的痕量铋、锑、硒和碲 被引量:2

Simultaneous determination of bismuth,antimony,selenium and tellurium in high-purity gold by hydride generation-four-channel non-dispersive atomic fluorescence spectrometry
在线阅读 下载PDF
导出
摘要 用乙酸乙酯萃取分离金,分液还原后应用氢化物发生-四通道原子荧光光谱法测定高纯金中的痕量铋、锑、硒和碲。在最佳条件下,方法对铋、锑、硒和碲的检出限分别为0.04,0.05,0.03,0.03 ng/mL(3δ);测定精密度(RSD)分别为0.98%,0.93%,0.94%,0.99%(对10 ng/mL Bi、Sb、Se和Te混合标准,n=7)。用所建立的方法对实际样品中的铋、锑、硒和碲进行了同时测定,测定结果与用标准方法测定所得结果之间无明显差异,各元素的加标回收率在95%~105%。结果表明,此方法能同时精确测定高纯金中的痕量铋、锑、硒和碲的量,且检出限较低,测定结果与国家标准方法一致。 Using ethyl acetate extraction and separation of gold,when the liquid after reduction,the simultaneous determination of bismuth,antimony,selenium and tellurium in high-purity gold sample using hydride generation-four-channel non-dispersive atomic fluorescence spectrometry has been developed.Under optimal conditions,the detection limits for Bi,Sb,Se and Te were determined to be 0.04,0.05,0.03,0.03 ng/mL(3δ),respectively.The precisions for 7 replicate determinations at the 10 ng/mL of Bi,Sb,Se and Te were 0.98%,0.93%,0.94% and 0.99%(RSD),respectively.The proposed method was successfully applied to the simultaneous determination of Bi,Sb,Se and Te in high-purity gold samples and compared with the standard methods,and recovery ratio of various elements in the range of 95%~105%.All the results of the present method were statistically equivalent to the results of the standard method.
出处 《应用化工》 CAS CSCD 2013年第8期1538-1540,1547,共4页 Applied Chemical Industry
关键词 高纯金 铋、锑、硒、碲 四通道无色散原子荧光光谱 同时测定 high-purity gold bismuth antimony selenium tellurium four-channel-nondispersive atomic fluorescence spectra simultaneous determination
  • 相关文献

参考文献8

  • 1中华人民共和国国家质量监督检验检疫总局,中国国家标准化管理委员会.GB/T25933—2010高纯金[S].北京:中国标准出版社,2011.
  • 2中华人民共和国国家质量监督检验检疫总局,中国国家标准化管理委员会.GB/T25934.1—2010高纯金化学分析方法第1部分:乙酸乙酯萃取分离-ICP-AES法测定杂质元素的含量[S].北京:中国标准出版社,2011.
  • 3中华人民共和国国家质量监督检验检疫总局,中国国家标准化管理委员会.GB/T25934.2—2010高纯金化学分析方法第2部分:ICP-MS-标准加入法校正-内标法测定杂质元素的含量[S].北京:中国标准出版社,2011.
  • 4Li Zhongxi,Yang Xiaoming,Guo Yuean,et al.Simultane-ous determination of arsenic,antimony,bismuth and mer-cury in geological materials by vapor generation-four-channel non-dispersive atomic fluorescence spectrometry[J].Talanta,2008,74:915-921.
  • 5倪迎瑞,李中玺,李海涛,闵丁丁.乙酸乙酯萃取-氢化物发生-原子荧光光谱法测定金锭中锑和铋[J].分析化学,2011,39(11):1774-1775. 被引量:13
  • 6Ni Yingrui,Li Zhongxi,Li Haitao,et al.Simultaneous de-termination of arsenic,antimony,selenium and telluriumin pure copper by hydride generation-four-channel non-dispersive atomic fluorescence spectrometry[J].Metallur-.
  • 7倪迎瑞,李中玺,李海涛,秦靖,张圣杰,薛振宇.化学蒸气发生-四通道原子荧光光谱法同时测定高纯金中的痕量As,Sb,Bi,Te[J].分析试验室,2012,31(7):58-60. 被引量:8
  • 8吴晓妮,倪迎瑞.盐酸沉淀分离-氢化物发生-四通道原子荧光光谱法同时测银锭中铋、锑、硒和碲[J].应用化工,2012,41(10):1823-1824. 被引量:3

二级参考文献14

  • 1GB/T41342003GoldIngots(金锭).NationalStandardsofthePeople’sRepublicofChina(中华人民共和围同家标准).
  • 2CB/T11066.82009MethodsJorChemicalAnalysisojGold—DeterminationofSilver,Copper,Iron,Lead,Anti—mony,Bismutk,Palladium,Magnesium,Tin,Nickel,ManganeseandChromiumContent—EthylAcetateExtraction—InductivelyCoupledPlasma—AtomicEmissionSpectrometry(金化学分析方法银、铜、铁、铅、锑、铋、钯、镁、锡、镍、锰和铬量的测定乙酸乙酯萃取电感耦合等离子体原子发射光谱法).NationalStandardsofthePeople‘sRepublicofChina(中华人民共和国国家标准).
  • 3I.iZX, YangX M, Guo Y A, Li H T, Feng Y H. Talanta, 2008, 74.. 915-921.
  • 4LiZX, Guo Y A. Talanta, 2005, 65:1318-1325.
  • 5GB/T25933-2010.高纯金.中华人民共和国国家标准.
  • 6GB/T25934.1-2010.高纯金化学分析方法第1部分:乙酸乙酯萃取分离-ICP-AES法测定杂质元素的,含量.中华人民共和国国家标准.
  • 7GB/T25934.2-2010.高纯金化学分析方法第2部分:ICP-MS一标准加入法校正一内标法测定杂质元素的含量.中华人民共和国国家标准.
  • 8Li Z X, Yang X M, Guo Y A, et al. Talanta, 2008, 74 : 915.
  • 9中华人民共和国国家质量监督检验检疫总局,中国国家标准化管理委员会.GB/T4135--2002银[s].北京:中国标准出版社,2002.
  • 10中华人民共和国国家质量监督检验检疫总局,中国国家标准化管理委员会.GB/T11067.3--2006银化学分析方法硒和碲量的测定电感耦合等离子体原子发射光谱法[s].北京:中国标准出版社,2007.