摘要 目的 评价国产XF 980 1型麻醉废气吸附器对氮氧化物、氟化物的吸附效果。方法 选择两个条件基本相同手术间 ,安氟醚或异氟醚 /氧化亚氮静吸复合麻醉的全麻病人 30例。在用国产XF 980 1型麻醉废气吸附器的前、后 ,对麻醉机废气排出口处氮氧化物及氟化物浓度进行监测。结果与使用麻醉废气吸附器前相比 ,使用麻醉废气吸附器后麻醉机废气排出口处氮氧化物、氟化物明显降低 (P <0 0 1)。结论 使用挥发性麻醉药时采用国产XF 980 1型麻醉废气吸附器可明显改善手术室内空气质量。 Objective To evaluate the effects of applying XF-9801 type of anesthetics waste gas absorber, an activated carbon filter,to removing nitrogen oxides (NOx) and fluoride. Methods Thirty patients, undergoing general anesthesia with enflurane or isoflurane/nitrous oxide, were enrolled in this study.The gas samples were taken at the waste gas outlet of anesthetic machine,to identify the concentrations of nitrogen oxides and fluoride before and after the application of anesthetics waste gas absorber. Results Compared with those before the absorber application, nitrogen oxides and fluroide concentrations in anesthetic waste gas decreased significantly following the absorber application (P<001). Conclusions The application of XF-9801 type of anesthetics waste gas absorber can greatly improve the air quality in operating room.
出处 《中华麻醉学杂志》 CAS CSCD 北大核心 2000年第9期524-525,共2页 Chinese Journal of Anesthesiology
关键词 麻醉药 清除 麻醉废气吸附器 氮氧化物 氟化物 Gas scavengers Anesthetics,inhalation Nitrogen oxides Fluorides
  • 相关文献

参考文献1

二级参考文献4

共引文献4

同被引文献21

  • 1王大波,姚卫兵.应用气相色谱仪测定活性碳吸附安氟醚废气的实验研究[J].中华麻醉学杂志,1993,13(1):57-59. 被引量:5
  • 2杨文卿,黄丽芳,林定.吸入全麻的废气污染与控制措施[J].国际医药卫生导报,2005,11(6):125-126. 被引量:9
  • 3张铁民,高风云,赵明新,牛润岭,吴志杰,雅健,白光昕,许考印.环路内活性炭过滤苏醒的临床应用[J].中华麻醉学杂志,1996,16(1):36-37. 被引量:2
  • 4陈渝,史长宁,张崇建.呼吸机应急调用的气源通用转接头制作与应用[J].医疗卫生装备,2006,27(7):79-79. 被引量:5
  • 5Dahm SL,Steptoe P,Luttropp HH.Reinstrup PCharcoal as an airwary isoflurane reflection filter.Eur J Anaesthesiol,1998,15(2):230
  • 6Imberti R,Preseglio A,Quiros E,et al.Low flow anaesthesia reduces occupational exposure to inhalation anaesthetics.Enviromentl and biological measurement in opiating room personnel.Acta Anaesthesiol Scand.1995,39(5) :586
  • 7Hojkjaer Larsen V,Severinsen I,Wasben J.Removal of halogenated anaesthetics from a closd circle system with a charcoal filter.Acta Anaesthesliol Scand,1989,33(5) :374
  • 8孙岩军 赵钢 李长杰.回路内麻醉气体吸附器对低流量麻醉苏醒时的影响[J].临床麻醉学杂志,2004,16(4):179-179.
  • 9Romano E,Auci A,Absorption of volatile anesthetics on activated charcoal.Efficiency of an experimsetal filter during low-flow circuit ventilation.Minerva Anestesiol.1995,61(10) :437
  • 10Baungarten RK.Siaple charcoal filter for closed circuit anesthesia.Anesthesiology.1985,63( 1 ):125