摘要 本试验应用补体结合试验方法对牛布鲁菌病进行检测,分别从溶血素效价测定、补体效价测定、抗原效价测定及血清效价测定4个方面分析了补体结合试验测量不确定度来源,并计算了合成不确定度、扩展不确定度,最终样品测量结果报告值为:X=(1.06±0.043)mg/mL(95%置信区间),即本试验中判定的可以抑制50%溶血的血清最高效价所对应的溶血素浓度为1.06mg/mL,不确定度U为0.043。 Bovine brucellosis was detected by complement fixation test using standard 96-well microti- tre plates with round (U) bottoms. The uncertainty origins of test results included hemolysin titer deter- mination, complement titer determination, antigen titer determination and test serum determination, re- spectively. Next, the combined standard uncertainty and the expanded uncertainty were calculated, and the final detection result of the test serum was. X= (1.06±0. 043)mg/mL(95%confidence interval). Herein,the hemoglobin concentration, which corresponds to the highest sera titer that can inhibit 50 % lysis of sensitized sheep red blood cells, is 1.06 mg/mL uncertainty U is 0. 043.
出处 《中国兽医杂志》 CAS 北大核心 2013年第5期65-69,共5页 Chinese Journal of Veterinary Medicine
基金 国家质量监督检验检疫总局课题(2010IK006)
关键词 牛布鲁菌病 补体结合试验 测量不确定度 Bovine brucellosis complement fixation test measurement uncertainty
  • 相关文献

参考文献8

  • 1国家认证认可监督管理委员会.SN/T1525-2005布氏杆菌病微量补体结合试验方法[S].2005-02-17.
  • 2中国合格评定国家认可委员会.实验室认可与管理基础知识[Z].2007.
  • 3赵凤华,任长生.移液器容量误差测量的不确定度评定[J].工业计量,2010,20(3):45-46. 被引量:2
  • 4秦霄雯,陈庆俊,任宏伟,崔涛.酶标分析仪吸光度测量结果不确定度分析[J].中国计量,2010(6):78-79. 被引量:1
  • 5National standard method: uncertainty meassurement in tes- ting(QSOP 4)[Z]. Standards Unit, Evaluations and Standards Laboratory Centre for infentions, Department for Evaluations, Standards and Training, Health Protection Agency of Ireland, 2008.
  • 6秦绪珍,高学慧,徐二木,赵芳,齐志宏,邱玲.ELISA方法检测CA242的不确定度评估[J].中国卫生检验杂志,2010,20(4):894-896. 被引量:6
  • 7中国合格评定国家认可委员会.CNAS-GL06:2006,化学分析中测量不确定度的评估指南[Z].
  • 8CNAAS-GL05:2011,测量不确定度要求的实施指南[Z].

二级参考文献8

  • 1林金丽.酶标法测定人体血清中乙型肝炎表面抗原的不确定度分析[J].中国卫生检验杂志,2005,15(5):625-626. 被引量:6
  • 2CNAS-CL02:2008.医学实验室质量和能力认可准则(ISO15189:2007)[S].
  • 3Cembal S, Ambrosio J, Armada C, et al. Intra - assay total uncertainty of results in immunoassay techniques [ J ]. J Immunoassay Immunochem,200d,25 ( 1 ) :1 - 15.
  • 4CNAS-GLD6:2006.化学分析中不确定度的评估指南[S].
  • 5Guide to the Expression of Uncertainty in Measurement (GUM) ; ISO Geneva: Schweiz, 1993; ISBN 92-67-10188-9.
  • 6Quantifying Uncertainty in Analytical Measurements, 2nd Edition, Eurochem/CITAC Guide (2000).
  • 7中国合格评定国家认可委员会.CNAS-GL05:2006测量不确定度要求的实施指南[S].
  • 8郭莉莉.酶标法检测人体血清中HIV-1/HIV-2型抗体结果不确定度评定[J].中国卫生检验杂志,2009,19(1):46-47. 被引量:6