多排螺旋CT低剂量扫描在成人颅脑检查中的应用价值 被引量:4

The Clinical Application of the Low-dose Multi-Slice spiral CT to Adult Cranial Examination
暂未订购
导出
摘要 目的评价多层螺旋CT适用于成人颅脑检查的低剂量扫描条件。方法将500名健康成年人随机分成5组:第1组采用常规剂量,第2~5组分别采用低剂量(200、160、100、80mAs)扫描。采用双盲法对图像质量独自评分,分析扫描剂量与图像质量评分的关系;比较第2~5组的CT剂量指数(CTDI vol)和平均剂量长度乘积(DLP)值相对第1组下降的比例。结果图像质量评分一致性好(κ=0.860,P=0.017),扫描剂量与图像质量主观评分呈正相关(r=0.512,P〈0.01),第1~4组图像质量差异无统计学意义(P〉0.05),而第5组图像质量明显下降(P〈0.05)。与第1组相比,第2~5组CTDI vol、DLP值均显著降低(P〈0.05)。结论管电流为100mAs时,多层螺旋CT既能保证成人颅脑检查的图像质量,又能显著降低辐射剂量。 Objective To analyze the clinical application of low-dose multi-slice spiral CT to adult cranial examination. Methods A total of 500 healthy adults received multi-slice spiral CT in our hospital were divided into 5 groups, which were group Ⅰ(receiving conventional dose), group Ⅱ(tube current was 200 mAs), group Ⅲ(tube current was 160 mAs), group Ⅳ(tube current was 100 mAs) and group Ⅴ(tube current was 80 mAs). The image quality was solely determined by double blind method and the relationship of radiation dose and image quality was analyzed. The reduction ratios of computed tomographic dose index (CTDI vol) and the average dose length product (DLP) in group Ⅱ to Ⅴ were compared to group Ⅰ. Results The concordance of the scores of image quality was perfect(κ=0.860,P=0.017). The positive relationship to radiation dose and image quality was confirmed. The image qualities of group Ⅰ to Ⅳ were no different (P〉0.05), but the image quality of group Ⅴ was poor (P〈0.05). Compared to group Ⅰ, the values of CTDI vol and DLP in group Ⅱ to Ⅴwere lower (P〈0.05). Conclusion The multi-slice spiral CT which the value of tube current was 100mA could ensure the image quality and reduce the radiation dose to adult cranial examination.
作者 范晓东
出处 《中外医疗》 2013年第10期3-4,共2页 China & Foreign Medical Treatment
关键词 多排螺旋CT 低剂量 体层摄影术 图像质量 辐射剂量 Multi-slice spiral CT Low-dose Tomography Image quality Radiation dose
  • 相关文献

参考文献7

  • 1Iball GR,Brettle DS,Moore AC.Assessment of tubecurrent modulation in pelvic CT[J].Bdt J Radiol,2006,79(937):62-70.
  • 2缪亚如,邵燕惠,薛跃君.螺旋CT肺部低剂量技术的研究[J].实用医技杂志,2011,18(10):1059-1061. 被引量:7
  • 3Stefania MK,Rizzo KM,Novelline RA.Reducing radiation dose in emergency computed tomography with automatic exposure control techniques[J].Emergency Radiology,2005,12(11):267-274.
  • 4华建明,陈根松,章伟敏.低剂量颞骨多层螺旋CT扫描技术的应用研究[J].中华放射医学与防护杂志,2006,26(5):530-532. 被引量:19
  • 5Menzel HG,Schibilla H,Teunen D.Guideline on radiation dose on the patient[J].European Guidelines on Quality Criteria for Computed Tomography,2006,113(13):1305-1310.
  • 6李坤成,王亮,王晓峰.CT机技术参数与CT片质量的相关性分析[J].中华放射学杂志,2004,38(4):373-376. 被引量:30
  • 7Kalra MK,Maller MM,Toth TL,et al.Techniques and applications of automatic tube current modulation[J].Radiology,2004,233(3):649-657.

二级参考文献18

共引文献53

同被引文献21

引证文献4

二级引证文献9

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部